P45000
27
Isolation of the variant with captive cable P45***K2*1*
Rated isolation voltage U
Nm
Reinforced insulation input against output/power supply
P450**K2*1*
1)
:
EN 50124-1 (rolling stock)
2300 V AC/DC
EN 50124-1 (fixed installations)
2300 V AC/DC
EN 50178
2300 V AC/DC
UL 347
2300 V AC/DC,
in approval
EN IEC 60664-1
1000 V AC / 1500 V DC
EN 61010-1
1000 V AC/DC
P451**K2*1*
1)
:
EN 50124-1 (rolling stock)
3600 V AC / 4800 V DC
EN 50124-1 (fixed installations)
3600 V AC / 4800 V DC
EN 50178
3600 V AC / 4800 V DC
UL 347
3600 V AC / 4800 V DC,
in approval
EN IEC 60664-1
1000 V AC / 1500 V DC
EN 61010-1
1000 V AC/DC
Function isolation input against input
EN 50124-1 (rolling stock)
3600 V AC / 4800 V DC
EN 50124-1 (stationary installations)
3600 V AC / 4800 V DC
EN 50178
3600 V AC / 4800 V DC
UL 347
3600 V AC / 4800 V DC,
in approval
EN IEC 60664-1
1000 V AC / 1500 V DC
EN 61010-1
1000 V AC/DC
Isolation inputs against environment
Measure distances to neighboring devices and conductive parts within the environment of the device according
to the applied standard. Carry out, evaluate and safeguard isolation coordination with clearance and creepage
distances (
Clearance and Creepage Distances, p. 28
) and the relevant standards (e.g., EN 50124-1).
Carry out cable laying according to EN 50343.
Isolation given by cable insulation with 3600 V AC/4800 V DC. Check whether additional isolation may be required.