E-118
ADJ Menu
Passcode Entry
Error
message
Prio-
rity
Error
Remedy
Er.06
1
EEPROM error
Send in the device
for repair
Er.10
4
Input connection does not cor-
respond to parameter setting.
Example: Input across
terminals 1 and 2 (4 - 20 mA)
and parameter setting for
0 - 20 mA
Change input
connection or
parameter settings.
Error
message
Prio-
rity
Error
Remedy
Er.07
3
Adjusting current unstable,
drift ≥ ±2 µA/s
Check loop current
and keep it constant
until adjustment is
terminated.
Er.08
3
Adjusting current invalid,
deviation at 4 mA > ±0.2 mA or
at 20 mA > ±0.4 mA
Check loop current
and remedy error.
Error
message
Prio-
rity
Error
Remedy
Er.09
3
Invalid passcode
Re-enter passcode.
If passcode is lost,
send in the device.
Содержание 830 R
Страница 2: ......
Страница 19: ...D 19 Deutsch Typschild Abb 5 6 Typschild Beispiel 1 Typschild außen 2 Typschild innen 3 Messwert zeichen 1 2 3 ...
Страница 25: ...D 25 Deutsch Messmodus Abwechselnd Messanfang und Messende anzeigen Schleifenstrom anzeigen ...
Страница 39: ...D 39 Deutsch S CON Menü für den min Kontakt beendet Parametrier Menü beenden ...
Страница 44: ...D 44 END anwählen und bestätigen zurück in Messmodus END ...
Страница 56: ...D 56 10 5 EU Konformitätserklärung 830 X S1 S2 R ...
Страница 57: ...D 57 Deutsch 10 6 EU Konformitätserklärung 830 S1 S2 R ...
Страница 72: ......
Страница 73: ...Process Indicator 830 X R S1 S2 Digital Indicators English E 3 www knick de S1 S2 R ...
Страница 74: ......
Страница 97: ...E 97 English Measuring Mode Toggle the display of start and end of scale Display the loop current ...
Страница 111: ...E 111 English End of S CON menu for min contact Exit parameter menu ...
Страница 116: ...E 116 Select END and confirm return to measuring mode END ...
Страница 128: ...E 128 10 5 EU Declaration of Conformity for 830 X S1 S2 R ...
Страница 129: ...E 129 English 10 6 EU Declaration of Conformity for 830 S1 S2 R ...
Страница 144: ......
Страница 145: ...Process Indicator 830 X R S1 S2 Afficheurs numériques Français F 3 www knick de S1 S2 R ...
Страница 146: ......
Страница 183: ...F 183 Français Menu S CON pour le contact min terminé Quitter le menu Paramétrage ...
Страница 188: ...F 188 Sélectionner et confirmer END pour retourner au mode Mesure END ...
Страница 200: ...F 200 10 5 Déclaration de conformité UE 830 X S1 S2 R ...
Страница 201: ...F 201 Français 10 6 Déclaration de conformité UE 830 S1 S2 R ...