4. Operation 

The pump must not start against pressure or vacuum. When it is 
switched on the pressure in the suction and pressure lines must be 
atmospheric. This also applies when the pump restarts after the 
power has been cut off for a short period.

 

The maximum permissible operating pressure (see section 10) 
must not be exceeded.

 

To prevent the maximum permissible operating pressure being 
exceeded, restriction or control of the air or gas flow should only be 
carried out in the suction line.

 

If restriction or control of the air or gas flow is made on the pres-
sure side ensure that the maximum permissible operating pressure 
is not exceeded. 

When the pump is at a standstill the inlet and exhaust must be at 
normal atmospheric pressure. 

Change the filter (accessories) if it is dirty.

 

The piston seals, and to a lesser extent the valve plates are the 
only parts of the pump subject to wear. Wear is usually indicated 
by a drastic reduction in the pneumatic performance. When   
replacing parts proceed as described in section 5. 

Ambient conditions: see section 1.3. 

Содержание UNPK09

Страница 1: ...ns The manual was produced for the serial pumps stated above With customer specified projects pump types starting with PU or MPU there could be differences in detail For customer specified projects please therefore take into account any agreed technical specifications as well as these instructions List of Contents Page 1 Description Operating Conditions 2 2 Safety 3 3 Installation 4 4 Operation 5 ...

Страница 2: ...sors and vacuum pumps must not be used for liquids You will find suitable liquid pumps in our Product Program For maximum permissible operating pressure ultimate vacuum and flow capacity see section 10 The KNF swing piston compressors and vacuum pumps must not be used in areas where there is a danger of explosion If your potential application lies outside the above limits discuss it with our techn...

Страница 3: ...upted by the thermal switch the pump will re start auto matically after cooling down Take all care necessary to prevent this leading to a dangerous situation Use only original KNF spare parts Pumps in the NPK 09 range conform to the safety regulations of the EC Low Voltage Directive 2006 95 EC and of the EC Directive 2004 108 EC concerning Electromagnetic Compatibility For the purposes of the Mach...

Страница 4: ...by more than 10 and 10 from that shown on the type plate The earth ground wire must be connected to the motor not necessary on dc motors up to 24 V For brushed dc motor lead wires connection Red V Black Ground Î Incorrect lead wire connection will damage motor For brushless dc motor lead wire connection Red V Black Ground White Voltage control 0 5 V DC 4 wire motor only Green Frequency output 4 wi...

Страница 5: ...n or control of the air or gas flow should only be carried out in the suction line If restriction or control of the air or gas flow is made on the pres sure side ensure that the maximum permissible operating pressure is not exceeded When the pump is at a standstill the inlet and exhaust must be at normal atmospheric pressure Change the filter accessories if it is dirty The piston seals and to a le...

Страница 6: ...r damage tongues bent corrosion The position numbers in the following text refer to figs 3 to 6 Fig 3 Pump element single head types Fig 4 Pump element twin head types Fig 5 Pump element all types Fig 6 Pump head all types Specifiaction Pos Description 1 Housing 2 Intermediate plate 3 Ribbed head 4 Screw 5 Connection only twin head models 6 Gasket 7 Valve Plate 8 Valve Plate 9 O Ring 10 Cylinder 1...

Страница 7: ...emove gasket 6 valve plate 7 intermediate plate 2 and valve plate 8 from housing 1 5 Only for the pump model NPK09 DC Take off the cover 18 after removing the 4 screws securing it For this pump type applies Where reference is made to turning or holding the cooling fan the necessary operations must be carried out by turning or holding the counter weight 17 6 Turn the fan to bring the retainer plate...

Страница 8: ...th valve plates are identical and they can be fitted either way up The intermediate plate can also be fitted either way up 6 Place the head assembly on the pump housing 1 as a guide to orientation see fig 3 single head models or fig 4 twin head models Reference point is the compression symbol indicated with a K in the diagrams or the mark M 7 Tighten the screws 4 at first lightly then firmly by ha...

Страница 9: ...using new valve plates 2 for each head For twin head pumps repeat the operations described for the second head and replace the connections between the two heads If after changing the valve plates the specified vacuum is not reached check the following pneumatic connections for tightness for twin head models that the head connections are gas tight If after changing the valve plates the pump produce...

Страница 10: ... inspect all parts for dirt before assembling the pump head and clean them if necessary As far as possible clean the parts with a dry cloth Solvents should not be used as they can attack the plastics and synthetic rubber parts If a compressed air line is available blow the parts out with it ...

Страница 11: ...ure or vacuum too low Compare the actual performance with the figures in section 10 or the data sheet There is pressure on the pressure side and at the same time vacuum or a pressure above atmospheric on the suction side The pump is not designed for this condition The internal diameter of connecting tubing or connected components or connected components is too small or they are restricted In order...

Страница 12: ...n side Connection for NPK 09 2 NPK 09 1 2 pressure side 026178 026179 6 Gasket 052650 7 8 Valve plate 052649 9 O Ring 025547 10 Cylinder 025528 11 Retainer Plate 025530 12 Support 025532 13 Piston seal 025531 14 Screw 110725 15 Washer 006886 Tab 6 Spare Parts according to figure 5 9 Accessories Description Order No Silencer Filter 072233 Hose Connector 1 8 MPTM HID 072235 Tab 7 Accessories ...

Страница 13: ...0 Tables Pump type Max permissible operating pres sure bar g Ultimate vacuum mbar abs Delivery rate l min at atm pressure UNPK09 7 100 12 UNPK09DC 7 100 15 UNPK09 1 100 24 UNPK09 2 7 24 UNPK09 1 2 7 100 24 Tab 8 Pneumatic Data Litre at STP ...

Страница 14: ...an RMA number submit a completed Decontamination Declaration form to rma knf com Î The Decontamination Declaration form can obtained from our website or by contacting KNF Technical Services http www knf com pdfs decontamdec doc Phone 609 890 8600 Î Product return instructions will be provided when the RMA is issued KNF Neuberger Inc 2 Black Forest Rd Trenton NJ 08691 1810 Phone 609 890 8600 Fax 60...