Installation Instructions UFK 1.1100 Ex
KNF Flodos UFK1.1100Ex O&M-Rev2
8
3 Installation
WARNING
Danger from incorrect installation
Injuries or damage to property from leaking media
After installation, check the system for leaks with a safe
medium.
Observe the requirements for intended use.
Take precautions in the design of the end device so that
leaking liquids cannot come into contact with live compo-
nents.
Use only connections that are specified in the data sheet.
Do not use NPT threads or sealing tape, as cracks can
occur at the plastic connections.
The following requirements apply to all activities described in this
section:
3.1 Installation location
The installation location must:
protect the pump against immersion
keep aggressive and flammable liquids and vapors away
allow hoses to be connected without tensile or bending stress
take account of cooling the pump drive
ensure that any leaks can be handled
have appropriate precautions if flammable media are used
3.2 Mechanical mounting
Protect the product against impacts, knocks, and strong vibra-
tions.
Please refer to the product data sheet for dimensions, installa-
tion positions, and fluid connections.
CAUTION
Danger due to loss of stability
Injuries caused by the pump toppling over
Wear safety shoes with steel caps.
When the pump is installed in the machine, secure it with
screws or clamps.
When attaching accessories, make sure that the pump can-
not topple over.