![KNF NF 1.60 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/knf/nf-1-60/nf-1-60_operating-and-installation-instructions_1980062016.webp)
Diaphragm-liquid pump NF 1.60
Servicing
KNF Flodos AG BA_NF1.60_EN_05_161566.docx
Translation of Original Operating and Installation Instructions
16
Mounting the pump head
13. Insert intermediate plate (
10
)
CAUTION: Make sure that the orientation is correct so that the
connections point in the right direction.
A pin and hole system prevents incorrect assembly.
14. Place the connecting plate (
7
) on top of the intermediate plate
(
10
).
15. Insert seal (
3
) and resonating diaphragm (
4
) in connecting
plate (
7
).
16. Place head plate (
2
) on connecting plate (
7
) and secure using
the 4 head screws (
1
).
17. Push hose with clamp ring (
6
) onto connecting plate (
7
) and
secure with union nut (
5
).
8.3. Replacing diaphragm, valves and seals
Same procedure as in Chapter 8.2, "Cleaning", except that the
rolled diaphragm (
11
), resonating diaphragm (
4
), anchor valves (
9
)
and seals (
8
) are replaced rather than cleaned.