Troubleshooting 

Diaphragm Liquid Pump NF 1.30 

 

KNF Flodos AG BA_NF1.30_EN_04_067674.docx 

18 

Translation of original Operating and Installation Instructions

 

 

Flow rate, suction head or pressure head is too low 

The pump does not achieve the performance stated in the technical data or on the data sheet. 
Cause 

Remedy 

Components in the system 
connected to the suction and 
pressure sides, such as hoses, 
valves or filters, are causing too 
much resistance 

  Modify installation, check the cross-sections of components 

Hose connections are not leak-
tight 

  Secure transition joints between hose and hose connector 

with clamps or clamping elements 

Particles in the pump 

  Clean the pump head; install suction-side filter if required 

(Spare Parts and Accessories), (see Chapter 8.2). 

Viscosity of the transferred 
medium is too high 

  Contact KNF 

Incorrect interchange of pres-
sure and suction line connec-
tions 

  Remove pressure and suction lines and reconnect correctly 

The pump parts are not re-
sistant to the medium to be 
transferred 

  Replace the pump head with a compatible version 

Tab. 12 

Fault cannot be rectified 

If you are unable to identify any of the above causes, please send 
the pump to KNF customer service (see address on first page). 

  Flush the pump to clear the pump head of any hazardous or 

aggressive liquids (see Section 8.2.1). 

  Dismantle the pump. 

  Clean the pump (see Section 8.2). 

  Send the pump, with completed decontamination declaration 

(see Chapter 11), to KNF customer service stating the nature 
of the transferred medium. 

Содержание NF 1.30

Страница 1: ...is a country specific designation with no technical relevance Contents Page 1 About this document 2 2 Use 3 3 Safety 4 4 Technical data 6 5 Assembly and function 8 6 Installation and connection 9 7 Op...

Страница 2: ...ntial danger It also indicates the possible consequences of failure to observe the warning The signal word e g Warn ing indicates the level of danger Here you will see actions for avoiding the danger...

Страница 3: ...he compatibility of the materials of the pump head pump housing diaphragm and valves with the medium The temperature of the medium must lie within the permissible temperature range see Chapter 4 The t...

Страница 4: ...and understood the Operat ing and Installation Instructions and in particular the Safety chapter Always ensure adherence to all pertinent accident prevention and safety regulations when working on and...

Страница 5: ...ted machinery may not be commissioned until such time as it has been determined that the machine in which the partly completed machinery is to be assem bled conforms to the provisions of the Machinery...

Страница 6: ...type KT stands for Assembly Material1 Pump head PP Valve socket PP FFKM Sealing washer FFKM Diaphragm PTFE Tab 2 1 according to DIN ISO 1629 and 1043 1 The pump head comprises an intermediate plate a...

Страница 7: ...head material and the hoses hose connectors used Hydraulic connections Parameter Value Recommended hose size ID mm OD mm 4 6 Tab 5 Specifications NF 1 30 DCG Motor voltage 12V 24V Power consumption W...

Страница 8: ...Working chamber 4 Diaphragm 5 Eccentric 6 Connecting rod 7 Pump drive Fig 2 Pump assembly The diaphragm liquid pumps are based on reciprocating displace ment pump technology The elastic diaphragm 4 i...

Страница 9: ...3 Mounting dimensions NF 1 30 DCG Make sure that the installation location is dry and the pump is protected against water in the form of rain spray splashes and drips Protect the pump against dust Pr...

Страница 10: ...red motor cable black motor cable 6 3 Hydraulic connection Only connect components to the pump that are designed to handle the hydraulic data of the pump see Chapter 4 Technical data Only use hoses th...

Страница 11: ...d contact with flammable materials WARNING Danger of bursting of the fluid system due to overpressure The pump builds up pressure A closed system can cause the max permitted operating pressure to be e...

Страница 12: ...and can also be regulated to some extent For details see Chapter 4 Technical data Duty cycle short cycle operation KNF pumps are designed for continuous operation Short start and stop cycles may adve...

Страница 13: ...e contact your local KNF partner for further in formation www knf com Switching off the pump KNF recommends if transferring aggressive liquids the pump should be flushed thoroughly prior to switch off...

Страница 14: ...gerous substances in the pump Depending on the medium transferred caustic burns or poisoning are possible Wear protective clothing if necessary e g pro tective gloves Flush the pump with a neutral liq...

Страница 15: ...ten it securely Press the edges of the diaphragm all around Ensure the sealing washers 4 are dust free and place them in the centre of the valve socket seat of the intermediate plate 6 one sealing was...

Страница 16: ...inal Operating and Installation Instructions CAUTION Escaping liquid After assembly the pump may not be leak tight due to incorrect assembly damaged or soiled seal faces or other reasons Run the pump...

Страница 17: ...ket and seals see Section 8 2 3 Replace pump Tab 10 Pump is not priming Cause Remedy Suction side of pump not con nected Connect the suction side of the pump Liquid in the container is too low Fill co...

Страница 18: ...mp Clean the pump head install suction side filter if required Spare Parts and Accessories see Chapter 8 2 Viscosity of the transferred medium is too high Contact KNF Incorrect interchange of pres sur...

Страница 19: ...sories Spare Parts Spare Part Item Number per pump Order No Delta PT screw 30x30 1 4 160308 FFKM sealing washer 4 4 060463 PP FFKM valve socket 5 2 060186 PVDF FFKM valve socket 5 2 060161 PTFE diaphr...

Страница 20: ...ions 11 Return of the pump KNF undertakes to repair the pump only under the condition that the customer provides a certificate of the pumped medium and cleaning of the pump For this purpose please fol...

Страница 21: ...Diaphragm Liquid Pump NF 1 30 For your notes KNF Flodos AG BA_NF1 30_EN_04_067674 docx Translation of original Operating and Installation Instructions 21...

Страница 22: ...KNF worldwide Please find your local KNF partners at www knf com...

Отзывы: