Diaphragm pump N96_AC
Operation
Translation of Original Operating and Installation Instruction, english, KNF 329772-329775 06/22
21
7 Operation
7.1 General
WARNING
Risk of burns from hot pump parts and/or hot
medium
Some pump parts may be hot during or after oper-
ation of the pump.
à
Allow the pump to cool after operation.
à
Take protective measures to protect against
touching hot parts.
CAUTION
Risk of injury from bursting hoses during pressure
applications due to excessively high temperatures
When operating the pump in pressure applica-
tions, hoses that are not designed for the head
temperatures of the pump at the respective oper-
ating point could become porous and burst.
à
Use temperature-resistant pressure hoses at
the pneumatic connections.
à
Wear protective equipment if necessary (e.g.,
safety gloves, hearing protection).
WARNING
Injury to eyes
Coming too close to the inlet/outlet of the pump
may result in injury to the eyes due to the present
vacuum/operating pressure.
à
Do not look into the pump inlet/outlet during
operation.
CAUTION
Risk of injury from freely rotating shaft end
Touching the pump at the end of the shaft may re-
sult in injury through burning and crushing.
à
Take protective measures to safeguard
against touching moving and hot parts.
à
Wear appropriate personal protective equip-
ment if necessary.
à
Only operate the pumps under the operating parameters and operating
conditions as described in Chapter
à
Ensure the proper use of the pumps (See Chapter Proper use).
à
Eliminate the possibility of improper use of the pumps (see Chapter Im-
proper use).
à
Observe the safety instructions (Chapter Safety).