Technical data
Diaphragm pump N96_AC
12
Translation of Original Operating and Installation Instruction, english, KNF 329772-329775 06/22
Electrical data
Parameter
Unit
Value
KN version
Value
KT version
Voltage
[V]
100 - 240
100 - 240
Frequency
[Hz]
50 / 60
50 / 60
Power P
1
[W]
25
21
Max. permissible
line voltage fluctu-
ations
[%]
± 10
± 10
Drive protection
class
IP20
IP20
Rated current con-
sumption
[A]
0.46 / 100V AC
0.24 / 240V AC
0.39 / 100V AC
0.23 / 240V AC
Tab.9 Electrical data
Weight
Pump type
Unit
Value
N96K_E-W
[kg]
0.7
Tab.10 Weight
Other parameters
Parameter
Unit
Value
Permissible ambient
temperature [°C]
[°C / °F]
+ 5 to see Fig. 1
0 if non-condensing (frost-free)
Permissible medium
temperature [°C]
[°C / °F]
See Fig. 2
Dimensions
See Dimensional drawing N96,
Chapter
Highest permissible rel-
ative air humidity of the
environment
80% for temperatures up to 31 °C,
decreasing linearly to 50% at 40 °C.
Maximum installation al-
titude [m above sea
level]
[m / ft
above sea
level]
2000
Gas tightness* of the
pump head (leak rate)
[mbar l/s]
< 6 x 10
-3
Protection class of
pump
IP20
Starts against
- Vacuum
- Pressure
KN
KT
100
2.5
130
2.5
Tab.11 Other parameters
*The gas tightness of the pump head is no longer ensured after the pump head is
opened or after replacing diaphragms and valve plates. A leak test can be used to
determine whether the original gas tightness is achieved again.