Installation and connection
Diaphragm Pump N936
14
Translation of Original operating and installation instruction, English, KNF 320276-320275 09/19
Installation location
à
Make sure that the installation location is dry and that the pump is pro-
tected against rain, splash, gushing and drip water as well as from
other contaminants.
à
Make sure that the installation location is accessible for service.
The IP protection class of the pump motor is specified on the type
plate.
à
Mount the pump at the highest point in the system to prevent conden-
sate from collecting in the pump head.
à
Protect pump from dust.
à
Protect pump from coating with greases and oils.
à
Protect pump from vibration and impact.
Installation position
à
The pump can be mounted in any installation position. Use metal
screws to fasten the pump at the indicated attachment points.
Fig. 4 Pump N936
To dissipate any condensate that may form, KNF recommends in-
stalling the pump at an incline (sloping from the suction side to the
pressure side, see Fig. 4).
Also observe the markings on the pump head here).
WARNING
Risk of injury during operation
à
Take protective measures against touching
the electrically live parts (electrical connection,
motor windings if applicable).
à
Take protective measures against touching
moving and hot parts.
à
Automatic restart after tripping of the self-re-
setting protection temperature limiter.
CAUTION
Risk of pump damage during operation
à
Take protective measures against the pene-
tration of foreign bodies into the pump.