Diaphragm Vacuum Pump N 828 / N 838
Operation
Translation of original Operating and Installation Instructions, english, KNF 121249-121517 09/19
37
The connection between the PC and motor controller can be oper-
ated as an RS-232 interface. Accordingly, in the operating system
it is managed as an additional COM connection and can be ad-
dressed with conventional terminal software.
Interface configuration
▪
Baud rate: 57600 bits/s
▪
Data bits: 8
▪
Parity: none
▪
Stop bits: 1
▪
Flow control: none
Tables Tab. 14 to Tab. 16 contain the necessary command sets,
shown as ASCII characters. When transmitting, the commands
must be followed by ASCII character <CR> (carriage return, deci-
mal value 013). The underlined expressions are not characters, but
symbols as explained in table Tab. 16.
Parameter
Command*
Function
Reply
Motor
dB
Start
S ; E
dE
Stop
S ; E
Speed
dS
nnnn
[
nnnn
= speed
value]
Set nominal
speed
ns ; E
Tab. 14: Control commands
* A pause of at least 25 ms is required after the “dB” instruction set.
Parameter
Command
Reply
Actual motor speed;
Operating current of
the motor;
Temperature of the
motor controller;
Fault status
pP
ni ; ii , ti , ei , E
Fault status (single
value)
gP
ei ; E
Nominal motor speed
gS
ns ; E
Minimum possible
motor speed
gSl
nl ; E
Maximum possible
motor speed
gSh
nh ; E
Software version
number
iV
V ; E
Tab. 15: Reed commands
Interface protocol