Diaphragm Vacuum Pump N 828 / N 838
Installation and connection
Translation of original Operating and Installation Instructions, english, KNF 121249-121517 09/19
23
WARNING
Danger of burns from hot surfaces
Hot surfaces may be caused by overheating of the
pump.
➔
Install the pump so that the motor fan can intake
sufficient cooling air.
➔
For pumps with fan: Install the pump so as accidental finger
contact with the fan is impossible.
➔
When installing, make sure that there are no combustible or
thermally malleable objects placed in the immediate ambient of
the hot pump parts (head, motor).
➔
Make sure that the installation location is dry and the pump is
protected against rain, splash, hose and drip water as well as
other pollutions.
➔
Make sure, that the installation location is accessible for
maintenance and service.
The IP protection class of the pump motor is indicated on the
type plate.
➔
Install the pump at the highest point in the system to prevent
condensate from collecting in the pump head.
➔
Protect the pump from dust.
➔
Protect the pump against grease and oils.
➔
Protect the pump from vibrations and jolts.
WARNING
Personal injury and/or damage to property because
of vibration
In conjunction with adjacent components, vibration of
the pump may result in crushing and/or damage to
these components.
➔
Make sure that vibrations of the pump do not
result in hazards associated with adjacent com-
ponents.
➔
Protect the pump against contact and intrusion of foreign mat-
ter.
Cooling air supply
Immediate ambient of the hot
pump parts
Installation location
Foreign matter protection