Diaphragm liquid pump with linear drive FMM 80
Installation and connection
KNF Flodos AG BA_FMM80_EN_08_164464.docx
Translation of original Operating and Installation Instructions
10
6.2. Electrical connection
Only have the pump connected by an authorized specialist.
Only have the pump connected when the power supply is
disconnected.
The electrical installation must be fitted with a device that
disconnects the magnetic coil from the mains (acc. to EN
60335-1)
When connecting the device to a power source, the relevant
norms, directives, regulations and technical standards must be
observed.
Connecting the pump
Make sure that the power supply data match the data on the
motor's type plate. The current consumption can be found on
the type plate.
The supply voltage must not deviate more than ±10% from the
specifications on the type plate.
Connect the voltage supply cables to the two pump leads.
The polarity is irrelevant.
CAUTION
Do not connect the pump to a continuous direct
voltage. Information on the control signal (see
Chapter 7
).
As a consequence of the built-in transient voltage
suppressor, the special requirements for the con-
trol electronics must be observed (see
Chapter
7.2
).