Installation Instructions FK 1100
Safety
KNF Flodos 178057_MA_FK1100_EN_00.docx
7
DANGER
Danger of injuries and damage to property due to leaks at the
interfaces to the pump head
Poisoning and chemical burns or unintended reactions caused by
leaking dangerous substances
➢
Wear personal protective equipment.
➢
Connect the pump correctly.
➢
Operate the pump only when you are sure it has no tech-
nical malfunctions.
➢
Operate the pump in accordance with the technical data.
DANGER
Danger of injuries and damage to property due to uncon-
trolled flow while the pump is not in operation
Poisoning and chemical burns or unintended reactions caused by
leaking dangerous substances
➢
Construct the fluid system so that the operating pressure at
the pressure side of the pump is higher than at the suction
side.
➢
Install a shut-off valve in the fluid circulation system.
The risk analysis indicates a risk of explosion from materials and
substances.
DANGER
Danger of injuries and damage to property due to an explo-
sion in the pump
The pump housing is made from aluminum. If the pump dia-
phragm is damaged, medium can accumulate in the housing and
form hydrogen (especially with acids and lyes). This could cause
an explosion in the pump housing.
➢
Check and monitor the reactivity of the medium with alu-
minum.
➢
No flammable materials in the direct vicinity of the pump or
piping
The pump housing is designed to withstand pressure surges
caused by explosions. The probability of a fault that would cause
an explosion is deemed to be low. When installed in the system,
the risks in the complete system must be assessed.