Technical Data
33
Technical Data
Wavelengths
190 - 740 nm,
∆λ
8 nm
Simult. Wavelengths
4 with digital control, 2 with usage of analog
outputs (,,stand alone“ mode)
Lamp
Deuterium (standard), Halogen (optionally)
Edge filter
370 nm
Wavelengths accuracy
± 1 nm
Scan function
on
fl
ow
Range of linearity
0-2.5 AU
Sensitivity
2 x 10
-5
AU at 240 nm and
time constant 1.0 s
Noise 1
λ
1 x 10
-5
AU at 240 nm, 1.0 s;
2
λ
2 x 10
-5
AU at 240/300nm,1.0 s
Baseline drift
1
λ
10 x 10
-5
AU/h at 240 nm, 1.0s;
2
λ
15 x 10
-5
AU/h at 240/300nm,1.0 s
Time constants
0.1/ 0.2 / 0.5 /1.0 / 2.0 / 5.0 / 10.0 s
Scalable
± 0,1 V / ± 1,0 V / ± 10 V
integrator out
16 steps
Autozero Range
Full Scale
Display
2 x 24 characters
Control
2 RS232 interfaces,
event outputs (TTL, OC, relays),
remote connectors, analog input
Miscellaneous
Storage of up to 20 programs,
Weight / Dimensions
5.5 kg, 160 x 185 x 240 mm (W x H x D)
GLP Support
Detailed trace report with operating hours of
total, lamp, servo motor; number of lamp
ignitions, comprehensive service information
Содержание WellChrom K-2600