23
Initial Startup
KNAUER
10D User Manual V7670, Version 2.3
Startup of the Device
CAUTION!
Possible damage to the device caused by con-
densed water!
Allow device to acclimate for 3 h, before it is con-
nected to power supply and taken into operation.
Prerequisite
Installation has been completed.
Flow cell has been assembled.
Procedure
1. Switch off the power supply.
2. Plug the power supply into the device.
3. Connect the power supply and the plug.
4. Plug in the power supply.
5. Using the power-supply switch, switch on the detector.
Result
The detector starts its self-calibration. The display shows the
progress that has been made. When the self-calibration has fin-
ished, the status display appears. If the self-calibration fails,
switch the detector off and back on.
Practical Tip
Regularly check the capillaries for leaks.