17
Unpacking and Setup
KNAUER
10D User Manual V7670, Version 2.3
Connecting Cables to the Pin Header
To control one device through another, you use the pin header.
The single ports are used to exchange control signals.
To use remote control, you have to connect cables to the pin
header. Therefore, connect the cables firstly to the spring strip
and push the spring contacts of the spring strip secondly onto
the pins of the pin header.
Prerequisites
The device has been turned off.
The power plug has been pulled.
Tools
Depressor tool
CAUTION!
Short-circuit hazard.
Turn off the device before connecting it to the
pin header. Pull the power plug.
CAUTION! Electrostatic discharge can destroy the
electronics!
Wear protective bracelet against electrostatic dis-
charge and ground.
ERROR IN/OUT
IN: After receiving a signal (short-cir-
cuit to GROUND) from an external
device, an error message appears and
the device stops.
OUT: An error signal is active for as
long as an error is displayed e.g. the
lamp does not start.
GROUND
Reference point for external analog
control of the detector.
ANALOG IN
Allows external analog control of the
detector when the option ANALOG
has been selected in the SETUP menu.
The control voltage must be applied
against GROUND.
Start IN
After receiving a signal (short-circuit to
GROUND) from an external device,
the device starts. If controlled via soft-
ware, an electronic trigger is sent via
LAN.
Connector
Function