AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
24
Product information
Plug
Function
START OUT
TTL output (default setting)/
OC with external pull-up to 24 V (25 mA)
Levels:
passive 5 V
active 0 V
START IN
TTL-compatible input
Low-active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to ground) from
an external device, the detector starts. If controlled
with software, an electronic trigger is send through the
LAN.
ERROR OUT
TTL output (default setting)/
OC with external pull-up to 24 V (25 mA)
Levels:
passive 5 V
active 0 V
Output remains active until the cause of the error has
been eliminated.
ERROR IN
TTL-compatible input
Low-active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to GND) from an
external device, an error message appears, and the
detector stops.
VALVE 24 V
Event-controlled switching of 24 V against GND.
5 V OUT
Provides a voltage of 5 V with respect to GND. This
makes it possible to supply a load that is switched by
an EVENT.
Protection: 5 V–50 mA
GND
Grounding of all inputs and outputs. Reference point
of the voltage at the signal inputs.
6.5.2 Connecting cables to the pin header
To control one device through another, the pin header is used. To use
remote control, you have to connect cables to the pin header. The single
ports are used to exchange control signals.
The device is switched off.
The power plug is disconnected from the device.
Operating tool
Prerequisites
Tools