AZURA® Detector RID 2.1L / RID 2.1L HighFlow Instructions V6750
16
Product information
OUT port
Procedure
Figure
1.
Push the capillary
1
through
the fitting
2
.
2.
Slide the lock ring
3
over the
tubing. Note that the tapered
end of the lock ring (with the
thin edge) has to point to the
seal ring
4
.
3.
Attach the seal ring
4
.
1
2
4
3
2
1
3
2
4
4.
Insert the capillary as far as
possible into the OUT port
5
at the front of the device.
5.
Tighten the fitting finger tight.
6.
Connect the capillary to the
OUT port
6
with the T-piece
7
and tighten the fitting finger
tight.
7.
Connect the backpressure reg-
ulator
8
between T-piece
and capillary
9
which leads
into a waste container. Note
the arrow on the backpressure
regulator which shows the flow
direction.
8.
Connect the capillary to the
fraction collection
bl
with the
T-piece and tighten the fitting
finger tight.
2
1
3
2
4
6
bl
8
9
7
Check if all fittings are tight.
6.3
Integrating the detector into the system
To integrate the detector into a system, note the ambient conditions
found in the section Technical Data as well as the ambient conditions of
other devices to be integrated into that system. The detector is integrated
into the HPLC flow system by connecting the capillary to the flow cell and
the HPLC system. The capillary connections in a basic HPLC system are
shown in the following picture.
5
Next Steps