 Installation

5

Airsensor Supplement, V6879

Installation

See below for a description on how to mount the airsensor onto the tub-
ing and on how to connect it to a computer. You can attach the airsen-
sor to the side panel of your AZURA L device with AZURA Click. Select a 
position for the air detection on the tubing, then start the installation.

NOTE

The number of the airsensor corresponds to the num-
ber on the cable that is being used. If you have more 
than one airsensor, connect the other amplifiers to the 
free cables.

Process

Figures

1. Connect the Sub-D15 plug 

1

 with the 

gameport adapter 

2

2. Connect the gameport adapter per USB 

to your computer.

3. Connect the amplifier 

3

 to one of the 4 

numbered cables.

4. Clamp the airsensor 

3

 onto the tubing 

at the selected position.

1

2

3

4

Содержание AZURA HPLC

Страница 1: ...HPLC Air sensor Supplement Document no V6879...

Страница 2: ...Phone 49 30 809727 0 Fax 49 30 8015010 Internet www knauer net E Mail info knauer net Version information Article number V6879 Version number 1 3 The information in this document is subject to change...

Страница 3: ...erve the warnings and safety information on the device and in the supplement NOTE Technical data are subject to change without notice Please check our website for latest updates and changes Intended U...

Страница 4: ...a computer The airsensor is controlled by KNAUER software PurityChrom which supports up to 4 airsensors Upon detecting air you can program differ ent actions The purification can either be stopped or...

Страница 5: ...detection on the tubing then start the installation NOTE The number of the airsensor corresponds to the num ber on the cable that is being used If you have more than one airsensor connect the other am...

Страница 6: ...y for approximately 3 s Integrating into PurityChrom You can program the software to either start or stop the system after an air bubble has been detected NOTE You have to choose the Gameport Input 1...

Страница 7: ...Process and figures 1 Open the PurityChrom Setup 2 Go to the register Communication 3 In the section Control Inputs you can make the necessary changes 4 To stop the system change the settings as follo...

Страница 8: ...ort Input 1 2 3 or 4 which equals the airsensor number Process Figures 1 To open the Time Control Editor press the button 2 Go to the register Threshold 3 Choose the Gameport Input 1 from the dropdown...

Страница 9: ...tivate the function Wait for Input Signal at a particular time during the method NOTE Make sure that the functions succeeding the function Wait for Input Signal start with a delay of 0 01 s NOTE Do no...

Страница 10: ...nsert to confirm the changes 7 After reaching the function Wait for Input Signal the run pauses until the end of sample After the signal the run contin ues automatically Name Order number Airsensor fo...

Страница 11: ...tions online https www knauer net en Support User manuals KNAUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 14163 Berlin Phone 49 30 809727 0 Fax 49 30 8015010 E Mail info knauer net Internet www kn...

Отзывы: