4. Installation and initial startup
Before you determine the operation site, read chapter 11 on page 43.
There you will find all device-specific information on power supply, am-
bient conditions and humidity.
Note:
The intended use be ensured only if the requirements for ambient
conditions of the operating environment are met.
4.1 Unpacking
Process
1.
Place the packaging in such a way that the lettering on the label is
in the correct position.
2.
Check the packaging, the device and the accessories for transport
damage.
3.
Check the scope of delivery. In the event of incomplete delivery,
contact Technical Support immediately.
4.
When lifting, carrying or moving the device, grab the unit only from
below on the sides. Do not hold onto front cover or leak tray, as
these parts are loosely attached to the device.
Keep the included packing list for repeat orders.
Keep the original packaging for safe storage or transportation.
4.2 Ambient conditions
4.2.1 Operation site
Observe the following requirements for the operation site so that the
measurement results are not influenced:
Place on a firm, level and straight surface.
Protect against direct sunlight.
Do not expose to air drafts such as air conditioning systems.
Do not set up the to other machines that cause floor vibrations.
Keep from high frequency sources.
Ensure adequate ventilation (see „4.2.3 Space requirements“, p. 10).
Avoid temperature fluctuations. (see „4.2.2 Ambient temperature“, p.
4.2.2 Ambient temperature
If the ambient temperature of the device is abruptly changed (e.g. when
it is installed in a cooling chamber), condensation will form inside the
device and may cause damage to the device. Allow the device to accli-
mate for 3 h, before it is connected to the power supply and taken into
operation.
Procedure
Next steps