 WARNINGS 

To ensure safe operation and service of the tester, follow these 

instructions. Failure to observe these warnings can result in severe 

injury or death.

•  LASER RADIATION

, Class II laser. 

DO NOT

 direct laser beam into eyes, 

DO NOT

 stare 

into the beam, or 

DO NOT

 view directly with optical instruments as this can cause 

severe and permanent eye damage.

•  Risk of electric shock and burn. Contact with live circuits could result in death or 

serious injury.

•  Use caution with voltages above 25V AC as a shock hazard may exist.

•  A blinking red or a steadily illuminated red NCV Indicator

 

5

 with audible beeps 

indicates the presence of voltage. A steadily illuminated blue NCV Indicator

 

5

 

indicates that no voltage is detected, however voltage could still be present.

•  Before and after each use, verify operation by testing a known working circuit that is 

within the rating of this unit.

•  Never assume neutral or ground wires are de-energized. Neutrals in multi-wire branch 

circuits may be energized when disconnected and must be retested before handling.

•  The tester 

WILL NOT

 detect voltage if:

•  The wire is shielded.

•  The operator is not grounded or is otherwise isolated from an effective earth ground.

•  The voltage is DC.

•  The tester 

MAY NOT 

detect voltage if:

•  The user is not holding the tester.

•  The user is insulated from the tester with a glove or other materials.

•  The wire is partially buried or in a grounded metal conduit.

•  The tester is at a distance from the voltage source.

•  The field created by the voltage source is being blocked, dampened, or otherwise 

interfered with.

•  The frequency of the voltage is not a perfect sine wave between 50 and 500Hz.

•  The tester is outside of operating conditions (listed in Specifications section).

•  Operation may be affected by differences in socket design and insulation thickness and type.

•  In bright light conditions, the LED visual indicators will be less visible.

•  When the NCV function is activated, the NCV Indicator

 

5

 will be illuminated either 

red or blue. DO NOT USE THE NVC FUNCTION UNLESS THE NCV INDICATOR

 

5

 IS 

ILLUMINATED.

•  Do not use if tester appears damaged or is not operating properly. If in doubt, replace 

the tester.

•  Do not apply more than the rated voltage as marked on the tester (1000V).

•  Detection within the range of 12V to 1000V is specified under “normal” conditions as 

detailed in the GENERAL SPECIFICATIONS section. The tester may detect at a different 

threshold at different conditions, or may not detect at all unless:

•  The tip of the tester is within 0.25" (6 mm) of an AC voltage source radiating unimpeded.

•  The user is holding the body of the tester with his or her bare hand.

•  The user is standing on or connected to earth ground.

•  The air humidity is nominal (50% relative humidity – non-condensing).

•  The tester is held still.

•  Always wear approved eye protection.

•  Comply with local and national safety requirements.

•  If this product is used in a manner not specified by the manufacturer, protection 

provided by the product may be affected.

ENGLISH

OPERATING INSTRUCTIONS

CHECKING FOR THE PRESENCE OF AC VOLTAGE

1.  Prior to use, test on a known live circuit to verify tester 

functionality.

2.  Place NCV Antenna 

7

 near an AC voltage source. If voltage is 

present, the unit will emit audible beeps and the NCV Indicator 

5

 

will illuminate.

SILENT MODE

With the NCVT-6 powered off, press and hold the On/Off Button for 

five seconds (the NCV On/Off 

1

 for NCV mode, or the LDM Control 

Button 

2

 for LDM mode) to power-on in Silent mode. While Silent 

mode is active, the silent mode icon 

H

 will be displayed. All visual 

indicators and display readings will continue to function as normal, 

but no audible indicators will be given. Silent mode will remain active 

until the NCVT-6 is powered-off. 

AUTO-POWER OFF (APO)

While the NCV function is active, the display 

4

 will powersdown 

following 15 seconds of no voltage detection (the NCV indicator 

5

  

will remain illuminated). After 4 minutes of further inactivity, the 

NCVT-6 will automatically power-off.
While the LDM function is active, the NCVT-6 will automatically 

power-off following 15 seconds of inactivity.

CLEANING

Be sure tester is turned off and wipe with a clean, dry lint-free 

cloth. 

Do not use abrasive cleaners or solvents.

STORAGE

Remove the batteries when not in use for a prolonged period 

of time. Do not expose to high temperatures or humidity. After 

a period of storage in extreme conditions exceeding the limits 

mentioned in the General Specifications section, allow the tester to 

return to normal operating conditions before using.

DISPOSAL / RECYCLE

Do not place equipment and its accessories in the trash. 

Items must be properly disposed of in accordance with local 

regulations. Please see 

www.epa.gov

 or 

www.erecycle.org

 

for additional information.

CUSTOMER SERVICE

KLEIN TOOLS, INC.

 

450 Bond Street, Lincolnshire, IL 60069

1-877-553-4676

[email protected]

www.kleintools.com

MAINTENANCE

BATTERY REPLACEMENT

When the Battery Status Indicator 

F

 shows only one bar remaining, 

the LDM functionality will be disabled, and the NCV functionality has less 

than 8 hours remaining. At this point, the batteries should be replaced:
1. Unscrew Battery Cap 

11

 and remove/recycle spent batteries 

10

.

2. Install two new AAA batteries 

10

. Note proper polarity.

3. Screw battery cap until fully seated to ensure a tight seal, being careful 

not to damage the O-ring 

9

NOTE: 

Damage to the O-ring can affect 

the IP40 dust resistant rating, but will not affect functionality. 

FUNCTION BUTTONS (FIG. 1)

NON-CONTACT VOLTAGE TESTING (NCV) ON/OFF BUTTON

 

1

To turn on the NCVT-6 and activate the Non-Contact Voltage Testing 

(NCV) function, press and release the NCV Power Button 

1

. The 

tester will emit a single audible beep and the NCV Indicator 

5

 will 

illuminate with a solid blue light. A proximity sensor indicating 

voltage strength using a dash scale will be shown on the LCD 

Display 

4

. Press and hold the NCV Power Button 

1

 for two 

seconds to power off the NCVT-6. 

NOTE:

 When the Laser Distance Meter (LDM) function is active, a 

short press of the NCV Power Button 

1

 will deactivate the LDM 

function and enter the NCV function.

LASER DISTANCE METER (LDM) ON/OFF BUTTON

 

2

To activate the LDM function, press the LDM Control Button 

2

• A short press of the LDM Control Button 

2

 will cause the NCVT-6 

to emit an audible beep and take a distance measurement. The value 

of the distance measurement will be shown on the Display 

4

. Press 

and hold the LDM Control Button 

2

 for two seconds to power off 

the Laser Distance Meter.

• A long press and hold of the LDM Control Button 

2

 will activate 

continuous measurement mode for the LDM function. As each 

new distance measurement is taken, the NCVT-6 will emit 

an audible beep and will update the display 

4

 with the new 

measurement value. 

WARNING: LASER RADIATION. DO NOT do any of the 

following, as severe and permanent eye damage could result:

• 

DO NOT

 direct laser beam into eyes.

• 

DO NOT

 stare into the beam.

• 

DO NOT

 view directly with optical instruments.

 CAUTION 

•  DO NOT

 attempt to repair this tester. It contains no serviceable parts.

•  DO NOT

 expose tester to extremes in temperature or high humidity.

GENERAL SPECIFICATIONS

The Klein Tools NCVT-6 is a full-range non-contact voltage tester  

with an integrated laser distance meter. It can detect voltage from 

12V to 1000V AC and can measure distance from 2 in. to 65.6 ft. 

(51 mm to 20 m). Under normal operating conditions, at distances 

from 51 mm (2 in) – 10 m (32.8 ft) the NCVT-6 has an accuracy 

of +/- 1.6 mm (1/16 in). At distances >10m (32.8 ft) – 20m (65.6 

ft ) the NCVT-6 has an accuracy of +/- 3.2 mm (1/8 in). The laser 

distance meter functions independently from the NCVT. 

•  Distance Measurement Range: 

2 in. to 65.6 ft. (51 mm to 20 m)

•  Voltage Detection Range: 

12V to 1000V AC

•  Frequency Range: 

50 to 500 Hz

•  Laser: 

Class II, 630 to 670nm, Max. Power <1mW

•  Batteries: 

2x AAA 1.5V Alkaline

•  Operating and Storage Altitude: 

Up to 6562 ft. (2000 m)

•  Operating and Storage Temp: 

32

°

 to 122

°

F (0

°

 to 50

°

C)

•  Relative Humidity:

 <85% non-condensing

•  Dimensions: 

6.45" x 1.32" x 1.21" (164 x 34 x 31 mm)

•  Weight: 

3.4 oz. (96 g) including batteries

•  Pollution degree: 

2

•  Safety Rating: 

CAT IV 1000V AC

•  Drop Protection: 

6.6 ft. (2 m)

•  Ingress Protection: 

IP40 Dust resistant

•  Standards:   Conforms  to

 UL STD. 61010-1, 61010-2-030 

Certified to

 CSA STD. C22.2 No. 61010-1, 61010-2-030

•  Complies with:

 21 CFR 1040.10 and 1040.11 except for deviations 

pursuant to laser notice No. 50, dated June 24, 2007.

Specifications subject to change.

FUNCTION BUTTONS (FIG. 1)

CHANGE UNITS / REFERENCE POINT BUTTON

 

3

• A short press of the Change Units / Reference Point Button 

3

 will  

switch the measurement reference point between Reference Point 1 

(NCV Antenna 

C

) and Reference point 2 (Battery Cap 

D

 ).

• A long press of the Change Units/Ref. Point Button 

3

 will change 

the units of measurement for the LDM function between meters, 

inches with decimals, inches with fractions, feet with decimals, and 

feet with fractions.

Содержание NCVT-6

Страница 1: ...ión ante caídas Protection contre les chutes 6 6 ft 2 m Safety Rating Clasificación de seguridad Cote de sécurité CAT IV 1000V Symbols on tester Símbolos del probador Symboles sur le testeur Warning Risk of electric shock Advertencia riesgo de choque eléctrico Avertissement Risque d électrocution Risk of danger Important information It is important that users of this tester read understand and fol...

Страница 2: ...ty After a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section allow the tester to return to normal operating conditions before using DISPOSAL RECYCLE Do not place equipment and its accessories in the trash Items must be properly disposed of in accordance with local regulations Please see www epa gov or www erecycle org for additional inform...

Страница 3: ...xponga a la humedad ni a altas temperaturas Luego de un período de almacenamiento en condiciones extremas que sobrepasen los límites mencionados en la sección Especificaciones generales deje que el probador vuelva a las condiciones de funcionamiento normales antes de utilizarlo ELIMINACIÓN RECICLAJE No arroje el equipo ni sus accesorios a la basura Los elementos se deben desechar correctamente de ...

Страница 4: ...ité élevés Après une période d entreposage dans des conditions extrêmes hors des limites mentionnées dans la section Caractéristiques générales laissez le testeur revenir à des conditions d utilisation normales avant de l utiliser MISE AU REBUT RECYCLAGE Ne mettez pas l appareil et ses accessoires au rebut Ces articles doivent être éliminés conformément aux règlements locaux Pour de plus amples re...

Отзывы: