ENGLISH
• TRMS
• AUTO / MANUAL
RANGE
• AUTO HOLD
• MAX / MIN
• BAR GRAPH
• BACKLIGHT
• LEAD HOLDER
• 3-5/6 DIGIT
6000 COUNT LCD
ENGLISH
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range
Resolution
Accuracy
600mV to 600V
0.1mV to 0.1V
± (0.5% + 4 digits)
1000V
1V
± (0.8% + 10 digits)
Overload Protection:
1000V
Input Impedance (Nominal)
: 10MΩ, <100pF
Input Impedance (Low Z)
: 3kΩ, <200pF, up to 600V RMS
AC Voltage Measurement
Range
Resolution
Accuracy
40Hz to 400Hz
Accuracy
400Hz to 5kHz
600mV
0.1mV
± (0.75% + 5 digits)
± (2.0% +3 digits)
6V to 600V
1mV to 0.1V
± (2.0% +3 digits)
1000V
1V
± (0.75% + 8 digits)
± (2.0% +3 digits)*
continued on next page.
Overload Protection:
1000V RMS
Input Impedance (Nominal)
:
10MΩ, <100pF
Input Impedance (Low Z)
: 3kΩ, <200pF, up to 600V RMS
Frequency Response:
40 to 5kHz
Response:
True RMS
*Accuracy for 400Hz to 1kHz range only.
DC Current Measurement
Range
Resolution
Accuracy
600µA to 400mA
0.1µA to 0.1mA
± (1.0% + 5 digits)
6A to 10A
1mA to 10mA
± (1.5% + 5 digits)
Overload Protection: µAmA:
F440mA/1000V fuse /
A Input:
F11A/1000V fuse
AC Current Measurement
Range
Resolution
Accuracy
40Hz to 2kHz
600µA to 400mA
0.1µA to 0.1mA
± (1.2% + 5 digits)
6A to 10A
1mA to 10mA
± (1.8% + 5 digits)
Overload Protection: µAmA:
F440mA/1000V fuse /
A Input:
F11A/1000V fuse
Frequency:
40 to 2kHz
Response:
True RMS
Resistance Measurement
Range
Resolution
Accuracy
600
Ω
to 6M
Ω
0.1
Ω
to 0.001M
Ω
± (1.0% + 5 digits)
60M
Ω
0.01M
Ω
± (1.5% + 10 digits)
Overload Protection:
600V RMS
Capacitance Measurement
Range
Resolution
Accuracy
60nF
0.01nF
± (3.5% + 6 digits)
600nF to 6000µF
0.1nF to 1µF
Overload Protection:
600V RMS
Frequency Measurement
Range
Resolution
Accuracy
99.99Hz to 399.9kHz
0.01Hz to 0.1kHz
± (0.1% + 3 digits)
Overload Protection:
600V RMS
Sensitivity:
> 2.5V RMS
Duty Cycle Measurement
Range
Resolution
Accuracy
1.0 to 99.0%
0.1%
± (0.2% per kHz + 0.1% + 5 digits)
Overload Protection:
600V RMS
Sensitivity:
> 2.5V RMS
Frequency Range:
<10kHz
Diode Test
Overload Protection
Range
Test Current
Open Circuit Voltage
600V RMS
6.0V
Appx. 0.25mA
< 1.8V DC (typ.)
Continuity Test
Overload Protection
Open Circuit Voltage
Threshold (Appx.)
600V RMS
< 0.44V
<40
Ω
Temperature Measurement
Range
Resolution
Accuracy
-50 to 600°C
0.1°C
± (1.0% + 3.0°C)
600 to 1000°C
1°C
± (1.0% + 3.0°C)
-58 to 600°F
0.1°F
± (1.0% + 5.4°F)
600 to 1832°F
1°F
± (1.0% + 5.4°F)
Overload Protection:
600V RMS
Thermocouple Accuracy:
Not specified
Specifications subject to change.
Do Not apply voltage to thermocouple.
• -50° to 1000°C (-58° to 1832°F)
8. Temperature:
Features:
Testing Fuse:
• When selecting uA or mA/A functions the meter will indicate
“
LEAd
” if the test leads are not properly connected to the meter.
• Note: This message remaining on the screen after properly
connecting the test leads is an indication that the fuse is open.
7.
Frequency (Hz) / Duty Cycle: <500kHz
Features:
HOLD
6.
Low Impedance (Low Z): < 600V
DO NOT use (LOW Z) mode at voltages greater than 600V.
• Press SELECT to choose AC or DC Voltage.
• Use Low Impedance (LOW Z) mode to reduce ghost voltages
• See AC/DC Voltage instruction
Capacitance Features:
HOLD
Safely discharge capacitor before
measurement.
• Reading may take up to 60 seconds
for large capacitors.
AUTO HOLD MAX/MIN
REL
HOLD
SYMBOLS USED ON LCD
AC Measurement
DC Measurement
-
Negative DC Value
AT
Auto Range Active
O.L.
Overload: Range Exceeded
Apo
Auto Power-Off Active
Low Battery
HOLD
Hold Active
MIN
Minimum Reading
MAX
Maximum Reading
%
Duty Cycle Mode
Frequency Mode
V
Voltage Measurement
A
Current in Amps
Ω
Resistance in Ohms
Diode Test
F
Capacitance in Farads
Continuity Test
Relative / Zero Mode
n
Nano 10
-9
µ
Micro 10
-6
m
Milli 10
-3
k
Kilo 10
3
M
Mega 10
6
°F
Degrees Fahrenheit
°C
Degrees Celsius
A-HOLD
Auto-Hold Active
LOW
Z
Low Impedance
WARRANTY
www.kleintools.com/warranty
CLEANING
Clean the instrument by using a damp cloth. Do not use abrasive cleaners
or solvents.
Take care to keep the sensor lens clean at all times. Clean the lens using a soft
cloth or cotton swab with water or rubbing alcohol only, and allow the lens to
dry before use.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
DISPOSAL / RECYCLE
Do not place equipment and its accessories in the trash. Items
must be properly disposed of in accordance with local regulations.
1000V
10A
PRESS