Kistler 8742A Series Скачать руководство пользователя страница 23

Shock Accelerometer 

Page 22 

 
002-096e-01.19 

 

 

8

 

Maintenance and Calibration 

8.1

 

General 

The  hermetically  sealed  construction  of  shock  accelerometers 
precludes  any  internal  maintenance  or  repairs.    External  cleaning  is 
important as covered in section 7.4. 

8.2

 

Trouble Shooting 

 

Should the user experience any problems with a shock accelerometer, 
the following procedure should be followed. 

Check  for  possible  intermittent  cables.    Faulty  cables  are  the  most 
common  cause  of  accelerometer  measurement  problems.    The 
application of a test signal at the accelerometer end of the cable is an 
effective means to determine if the problem is accelerometer, cable, or 
measurement instrument related.   

Kistler couplers are equipped with a bias monitor which can be used to 
check  the  cable  integrity  and  detect  malfunctions  in  the 
accelerometer.  Use the bias monitor as follows: 

 

 

A.  With the accelerometer connected, check for a normal reading on  
 

the bias indicator (normal should be around 11 VDC). 

B.  If the bias reading is high: 

1.  Disconnect  the  accelerometer  at  the  accelerometer  end  of  
 

the cable. 

2.  If  the  bias  indicator  reading  decreases;  the  accelerometer  is  
 

defective. 

C.  If the bias indicator reads low, disconnect the accelerometer  

1.  If the bias indicator still reads low, replace the cable. 

2.  If the bias level changes when the cable is disconnected, the  
 

accelerometer is defective. 

Analysis  and  measurement  equipment  with  built-in  accelerometer 
sources generally do not have bias monitors.  In these cases, the user 
can attach a T connector to the measurement/analysis instrument and 
measure the actual bias voltage.  If the system is operating properly, 
the  bias  voltage  should  measure  approximately  11  Volts.    Defects 
should show-up as in A and B above. 

If  it  is  determined  that  the  accelerometer  requires  repair,  please 
contact  the  factory  or  your  local  distributor/representative  for  repair 
information. 

 

 

Содержание 8742A Series

Страница 1: ...002 096e 01 19 Instruction Manual Shock Accelerometers Type 8742A...

Страница 2: ...Shock Accelerometer Page 1 002 096e 01 19 Instruction Manual Shock Accelerometers Type 8742A CE Kistler Instrument 75 John Glenn Drive Amherst NY 14228 2171 Tel 716 691 5100...

Страница 3: ...curing Cables 9 6 Operation 10 6 1 Powering 10 6 1 1 Using Built In Power Sources 10 6 1 2 Kistler Couplers 11 6 1 3 The Constant Current Power Supply Coupler 11 6 1 4 Sensor Power Up 14 6 1 5 Overloa...

Страница 4: ...instructions carefully before installing and operating these instruments Kistler offers a wide selection of measuring instruments and comprehensive systems including Sensors for force pressure accele...

Страница 5: ...ry area Keep the connector clean and covered when not in use Exercise care during cable connection Cable connector center pin should align with the female receptacle on the accelerometer Follow the re...

Страница 6: ...ccelerometers are shear mode shock and vibration measuring instruments These units are intended for pyroshock testing metal to metal impact studies drop testing etc A self generating piezoelectric sen...

Страница 7: ...material to produce an electric charge Since force is mass times acceleration from Newton s second law the charge produced is proportional to acceleration since m is constant This is represented by T...

Страница 8: ...r 4 Cable 5 Readout or analysis equipment 6 Installation 6 1 General For proper operation of shock accelerometers care must be taken during the installation process Careful installation will result in...

Страница 9: ...re that the base of the accelerometer makes full contact with the mounting surface See table 3 2 Completely clean the surface prior to mounting 3 Apply a thin coat of silicon grease to both the accele...

Страница 10: ...1 2 minimum 28 UNF 0 22 5 6 0 42 10 7 Table 3 Mounting Surface Preparation 6 4 Securing Cables Figures 4 and 5 show the correct and incorrect ways for installing cables Allow a sufficient radius to en...

Страница 11: ...7 1 1 Using Built In Power Sources Many FFT analyzers and vibration monitors are available with internal accelerometer power supplies Often identified as Constant Current Sources these internal power...

Страница 12: ...Type 5114 plus front panel adjustable gain and filter provision 5148 16 channel rack mountable AC powered coupler with unity gain LEDs indicate circuit integrity Front panel BNC output connectors 5165...

Страница 13: ...7 shows a schematic of a simple power supply coupler for shock accelerometers DC power is supplied from a 24 volt source such as a regulated power supply or batteries One advantage of the Piezotron s...

Страница 14: ...Es 1 Eb Maximum negative amplitude Fs Eb 2 Where Fs Full scale output peak ES DC supply voltage sometimes called compliance voltage Eb bias voltage from calibration certificate Figure 8 Output Signal...

Страница 15: ...ing range certain components will become non operational During this non operational state the amplifier components are protected from overload damage The amount of time required for recovery from an...

Страница 16: ...he cable capacitance becomes significant thereby loading the Piezotron impedance converter If the current is not sufficient to charge the cable at an adequate rate high frequency distortion will be ex...

Страница 17: ...creased the noise produced by the Piezotron circuit will also increase The miniature Piezotron amplifier will produce more heat as current is increased resulting in reduced life and reliability The sp...

Страница 18: ...ed reference frequency over a specified bandwidth Shock accelerometers are specified with maximum variations of 5 to 10 Kistler uses the industry standard reference frequency of 100Hz Many users may b...

Страница 19: ...ition will occur per the above equation which causes the sensitivity to increase with frequency It is desirable to have a uniform frequency response over the specified frequency range As can be seen F...

Страница 20: ...cer The low frequency response of accelerometers is an electrical limitation The design of shock accelerometers is such that low frequency drift caused by temperature variations is virtually eliminate...

Страница 21: ...ncy is the reciprocal of time The low frequency limit of shock accelerometers is defined as the point where the response is 5 also 10 where specified below the reference point 100 Hz The low frequency...

Страница 22: ...by providing optimal low frequency response When using a simple coupler circuit Figure 7 with AC coupling a time constant equal to the product of the decoupling capacitor s C capacitance times the rea...

Страница 23: ...onitor as follows A With the accelerometer connected check for a normal reading on the bias indicator normal should be around 11 VDC B If the bias reading is high 1 Disconnect the accelerometer at the...

Страница 24: ...cturer s product data for recommended removal agent Do not us an abrasive on the base surface This can affect the flatness thereby reducing high frequency transmissibility 8 5 Calibration Calibration...

Страница 25: ...Premium industrial cable connector 10 32 pos to BNC pos 1761sp x General purpose cable connector 10 32 pos to BNC pos 1762sp x General purpose cable connector 10 32 pos to BNC pos 1601Bsp BNC pos to...

Отзывы: