3.16.7.5 Azimuth Fine
Important: Before starting the alignment please verify the following conditions:
- RaZON
+
needs to be supplied with power (24 VDC -16%, +25%)
- RaZON
+
is sun tracking (not demo mode)
- Sun with unobstructed sky (DNI >300 W/m
2
) is required
- The RaZON
+
is correctly levelled (RaZON
+
bubble level in center)
- E-marking on flange is pointing East
- The 3 M4 bolts in the tripod / pole mount are fastened
When the above requirements are met, the final adjustment can be made by rotating the RaZON
+
with the buttons on the screen
until the sun falling through the first alignment target hole from the pyrheliometer makes a round spot of light around the second
alignment target hole. It takes about 10 seconds for the RaZON
+
to react on the left/right buttons. Confirm the correct setting with
[Done √]. The vertical adjustment of the side mounting plate is a factory setting and should be OK.
Done
Содержание RaZON+
Страница 1: ...RaZON Solar Monitoring System Instruction Manual ...
Страница 2: ...2 ...
Страница 4: ...4 ...
Страница 6: ...6 ...
Страница 12: ...12 ...
Страница 42: ...42 ...
Страница 44: ...44 ...
Страница 50: ...50 ...
Страница 52: ...52 ...
Страница 71: ...71 Example of a logged ASCII data file via Wi Fi Logging Eport Log File ...