Step 4: Add the 3 levelling screws (e) to the top flange of the pole mount. There is thread in the top flange and screwing the adjustment
screws in through the flange will raise the RaZON
+
on that side. Enabling the levelling of the RaZON
+
.
Only after finishing the levelling and azimuth adjustment the RaZON
+
can be secured fully by tightening the screws (d) from step 3.
When different mounting screws (4) are used they must not protrude through the bottom flange of the RaZON
+
When installed an initial Azimuth orientation due East with the help of a compass (smartphone) will do. When the RaZON
+
is powered
later on, exact positioning can be performed via Ethernet or Wi-Fi, when sun is available.