21. Enter Setup Mode:
21.1 If need to set configuration parameters, set user parameters, calibrate the scale, set current date or time,
test some hardware… It’s need to long press
SETUP
key to enter setup mode
21.2 After Entering Setup Mode, the main menu item
CONFIG
will be shown first.
21.3 In Setup mode, use to select wanted submenu and wanted
menu item, select wanted choice, set wanted number, confirm and save data, or exit this mode. Please
refer the
“4.Operation Menu Structure”
.
22. UPS world ship:
22.1. To interface the scale with UPS world ship set the following parameters:
Enter USER menu, set following parameters, if RS232 is used, set these on COM1, if USB is used, set
these on COM2:
USER-COM1(or COM2)-BAUD.RT = 4800
USER-COM1(or COM2)-BYT.FMT = 7E1
USER-COM1(or COM2)-LAYOUT = SCP-12
22.2. Set scale port to NCI3835 in UPS WorldShip:
23. Error codes and Troubleshooting
25.1 Meaning of Symbols:
25.1.1
0
﹉ ﹉ ﹉ ﹉
------ Zero is over the setting range
25.1.2
0
﹍ ﹍ ﹍ ﹍
------ Zero point is below the setting range
25.1.3
Ad
﹉ ﹉ ﹉
------ Signal to ADC is over max. range
)
25.1.4
Ad
﹍ ﹍ ﹍
------ Signal to ADC is below min. range
25.1.5
﹉ ﹉ ﹉ ﹉
------ Weight is over upper limitation, or display data is over limitation
25.1.6
﹍ ﹍ ﹍ ﹍
------ Weight is below lower limitation
25.1.7
EEP.E1
------ CONFIG or CAL parameters are not correctly set
25.1.8
EEP.E2
------ USER parameter is not correctly set
25.1.9
Lo.bAt
------ Battery voltage is lower than setting.
25.1.10
CAP.- - -
---- Next displaying content is Capacity
25.1.11
CAP.ER
----- Parameters about Capacity is not correct
25.1.12
CAL.Px
----- Calibration on point(x)
25.1.13
CAL.OFF
----- Calibration Seal Switch is on OFF position
25.1.14
CAL.ON
----- Calibration Seal Switch is on ON position
25.1.15
CAL.Er
----- Calibration error, maybe input data or loaded weight is incorrect, unstable, un-linear
25.1.16
CAL.End
----- Calibration is end
25.1.17
OFF
----- Power OFF the indicator
25.1.18
STB.ER
------ Unstable time is lager than setting of USER-OTHER-NLD.RNG
25.1.19
AC.xxxx
----- Accumulation times is xxxx