WARNINGS AND CAUTIONS FOR OPERATING THE TESTER

 7

 

2.2 Warnings and Cautions when Operating the 

Tester

 

2.2.1 Connecting of test lead-wire on LOW side

 

Fig.2-2 shows the connection of the test lead-wire on LOW side.  Every time the  Tester is used,
check if the lead wire is not damaged or disconnected.  The lead-wire connection to the DUT shall
be made from LOW side.  If the connection is incomplete, it is hazardous that the entire DUT may
be changed with high voltage.

 

Fig.2-2

Connecting of test lead-wire on LOW side

 

2.2.2 Connection of test lead-wire on high-voltage output side

 

After connecting the lead-wire on LOW side, take the following procedure.

 

1.

Push the STOP switch.

2.

Con

fi

rm if the indicator of the output voltmeter is at "0".

3.

Con

fi

rm if the DANGER lamp has been off.

4.

Connect the high-voltage test lead-wire with the HIGH VOLTAGE terminal. 

5.

Short the high-voltage test lead-wire with the LOW test lead-wire once, and con-

fi

rm that no high voltage is output.

6.

Lastly, connect the LOW test lead-wire and then the high-voltage output test lead-
wire to the DUT.

 

2.2.3 Power switch

 

The power switch shall be turned on after con

fi

rming that the TEST VOLTAGE dial has been turned

to the extreme left ("0" position).

 

2.2.4 Change of test condition

 

Change-over of the switches on the panel shall be made after con

fi

rming that the STOP switch has

been pushed in and that the TEST VOLTAGE dial has been turned to the extreme left ("0" position).

LOW terminal

Test lead-wire (Black) to housing of 

Device Under Test

Attach the test lead-wire fixuture, 

as illustrated.

Содержание TOS8870A

Страница 1: ...RESISTANCE TESTER OPERATION MANUAL DANGER This Tester generates high voltage Any incorrect handling may cause death Read Chapter 2 WARNINGS AND CAUTIONS FOR OPERATING THE TESTER in this manual to prev...

Страница 2: ...ets lost or soiled a new copy can be provided for a fee In either case please contact Kikusui distributor agent and provide the Kikusui Part No given on the cover This manual has been prepared with th...

Страница 3: ...eath by electric shock the output terminal the test lead wire connected to the output terminal the Device Under Test DUT any part of the Tester which is electrically connected to the output terminal a...

Страница 4: ...haracteristics suitable for this product The use of a fuse with a different rating or one that short circuits the fuse holder may result in fire electric shock or irreparable damage AC power cord The...

Страница 5: ...me or consumer use Purposes of use Do not use the product for purposes other than those described in the operation manual Input power Use the product with the specified input power voltage For applyin...

Страница 6: ...l be included when the product is relocated Operation Check that the AC input voltage setting and the fuse rating are satisfied and that there is no abnormality on the surface of the AC power cord Be...

Страница 7: ...IN JAPAN WARNING TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING CONDUCTOR OR THE PROTECTIVE CONDUCTOR TERMINAL MUST BE CONNECTED TO GROUND DO NOT REMOVE COVERS REFER SERVICING TO QUALIFI...

Страница 8: ...ich if ignored may result in damage to the product and other property Shows that the act indicated is prohibited Is placed before the sign DANGER WARNING or CAU TION to emphasize these When this symbo...

Страница 9: ...p plied from the fixed installation Examples of such equipment are appliances portable tools and other household and similar loads If such equipment is subjected to special requirements with regard to...

Страница 10: ...rnings and Cautions when Operating the Tester 7 2 2 1 Connecting of test lead wire on LOW side 7 2 2 2 Connection of test lead wire on high voltage output side 7 2 2 3 Power switch 7 2 2 4 Change of t...

Страница 11: ...ting the LOWER REFERENCE value 15 3 3 Single Insulation Resistance Test 17 1 Selecting the insulation resistance test mode 17 2 Check the Zero ohm 17 3 Selecting a test voltage 17 4 Setting the PASS F...

Страница 12: ...n of Rear Panel Items 35 Chapter 6 MAINTENANCE AND CALIBRATION 6 1 Cleaning the Outer Surfaces 36 6 2 Inspection 36 6 3 Maintenance 36 6 4 Calibration 36 Chapter 7 SPECIFICATIONS 7 1 Withstanding Volt...

Страница 13: ...reset on the front panel of the Tester but also when it is less than the low limit reference value which is continuously adjustable down to one half of the high limit reference value thereby contribut...

Страница 14: ...able atmosphere To prevent explosion or fire do not use the unit near alcohol or thinner or in an atmosphere con taining such vapors Avoid locations where the unit is exposed to high temperature or di...

Страница 15: ...ting in electrical shock or fire Do not use the unit in locations where highly sensitive measuring instru ments or receivers are nearby Such instruments may be affected by the noise generated by the u...

Страница 16: ...s posted near the AC line power input terminal of the Tester If no particular indication is made on this indicator sheet the AC line setting of the Tester is for the standard nominal voltage 100 VAC A...

Страница 17: ...panel 4 Plug in the AC power cord 1 6 Grounding Electric shock may occur if proper grounding is not furnished Connect the ground terminal to an electrical ground safety ground If you do not ground the...

Страница 18: ...nded when the Tester output is shorted to an earth line or to a conveyor or other device which is connected to an earth line or when it is shorted to the AC line the Tester chassis can be charged up t...

Страница 19: ...h 2 Confirm if the indicator of the output voltmeter is at 0 3 Confirm if the DANGER lamp has been off 4 Connect the high voltage test lead wire with the HIGH VOLTAGE terminal 5 Short the high voltage...

Страница 20: ...d the high voltage areas test lead wire probe or output terminal for correction of the wiring or any other purpose provided that the following confirmation has been made the indication of output volta...

Страница 21: ...nnect the DUT from the Tester during testing If it is disconnected check the two points above before touching high voltage areas 2 Consideration of discharge time The time required to discharge electr...

Страница 22: ...t short circuit the Tester output with the ground or a conveyer or any device connected to the ground or with the commercial power line around Tester location Such a short circuit may cause high volta...

Страница 23: ...ted at the CUTOFF CURRENT of 100 mA leave a sus pension time longer than the test duration Incidentally the maximum allowable test duration is 30 minutes at ambient temperature of 40 C or lower The ab...

Страница 24: ...ART switch The conditions for illuminating of the lamp are given on 5 1 Descrip tion of Front Panel Items on page 31 Eliminate the conditions and then press the STOP switch to reset the Tester When th...

Страница 25: ...page 19 5 Setting the test voltage 1 Set both TIMER ON OFF switch and LOWER ON OFF switch to the OFF state 2 Check that the PROTECTION lamp is not illuminating If the PROTECTION lamp is illuminating p...

Страница 26: ...luminate to indicate that the test volt age is ready to be applied to the DUT 5 Gradually turn clockwise the TEST VOLTAGE dial until the necessary test voltage is obtained 6 When the required test tim...

Страница 27: ...t is possible that the cur rent which flows through stray capacitances of the test leadwires and other DUT becomes larger than the low limit reference value and no low limit judgement for low leakage...

Страница 28: ...e when the leakage current of the DUT is smaller than the minimum detectable current of the Tester or when the current which flows in the stray cappacitances of the test leadwires and other device tes...

Страница 29: ...stances The window comparator system is especially advantageous when the distribution of the insu lation resistances of DUT can be predicted and the predicted high limit value is in the range of the d...

Страница 30: ...of the time setup switch equals the timer setting When the range selector switch is pressed the timer setting equals the setting of the time setup switch multiplied by 0 1 The setting of the timer se...

Страница 31: ...test If the Tester is in the RESET state the test will start as you press the START switch As you press the STOP switch the output voltage is cutoff and the test ends 9 Reapplication of test voltage r...

Страница 32: ...to the DUT 7 Connect the high voltage test leadwire to the DUT 4 Test procedure 1 Set the Tester in the RESET state 2 Press the START switch 3 The Tester will automatically perform the two types of t...

Страница 33: ...that the high test voltage is not generated inadvertently Also provide full measures to ensure that nobody is con tacted with the high voltage areas DUT test lead wire probe or output terminal etc whe...

Страница 34: ...er A timing of STOP signal for clearing of FAIL alarm sends STOP signal more than 20 ms after doing 100 ms or more after FAIL signal generated as follows Fig 3 7 With a view to prevent erroneous opera...

Страница 35: ...e interlock terminal is short circuited with a shorting piece to cancel the protection mode on a casual basis To perform a with standing voltage test you must use the interlock function to operate the...

Страница 36: ...conditions are met Power of the Tester is turned on The STOP signal is not applied or the STOP switch is not pressed Not in the TEST ON state Not generating the PASS signal Not generating the FAIL al...

Страница 37: ...en by this signal The minimum rating of the load applicable to the contact is 5 V 10 mA or higher Set V and R so that a load with higher rating is applicable Noise measures are also required in some c...

Страница 38: ...you press the START switch the Tester is driven into the TEST ON state In this mode testing cannot be performed simply by pressing the START switch Testing will only start when the STOP switch is pres...

Страница 39: ...perating the Tester in the MOMENTARY mode and using RC02 TOS Remote Control Box both hand type option in conjunction When test is made with the timer the Tester generates a PASS signal for approximate...

Страница 40: ...CONTROL CIRCUIT FAIL ALARM START STOP PROTECTION CIRCUIT HIGH LIMIT REFERENCE VALUE SETTING CIRCUIT LOW LIMIT REFERENCE VALUE SETTING CIRCUIT CURRENT DETECTOR CIRCUIT WITH LOG CONVERTER 1 2 4 3 5 6 7...

Страница 41: ...tisfy the specification accuracy 5 Ohmmeter This meter is driven by the above logarithmic signal to indicate the insulation resistance of the DUT 6 High limit reference value setting circuit Sets the...

Страница 42: ...alue set with the LOWER REFERENCE value control When the leakage current is not within the above range the comparator generates a FAIL judgement 4 3 Zero Start Switch If a regular mechanical contact t...

Страница 43: ...h is pressed the ohmmeter indicates the value of measuring resistance When the LOWER switch is pressed the ohmmeter indicates the low limit reference value for PASS FAIL judgment when UPPER switch is...

Страница 44: ...leakage current is more than the high limit reference value The LOWER ON OFF switch enables or disables this function 10 LOWER REFERENCE The potentiometer is used to set the low limit reference value...

Страница 45: ...illuminates When this has occurred remove the cause of the trip and then press the STOP switch to reset the Tester 1 When the RANGE 2 5kV 5kV switches of withstanding voltage tester are changed 2 Whe...

Страница 46: ...lecting the timer range on page 13 19 TEST VOLTAGE The dial controls the output voltage for withstanding voltage test The output voltage increases as you turn this dial clockwise Be sure to set the di...

Страница 47: ...milliammeter between them The milli amme ter should be capable of measuring the leakage current Be sure to connect the shorting bar when the current is not monitored A current may flow also when in in...

Страница 48: ...cuited by using a circuit tester or other instrument If the test leadwire is damaged its sheath is damaged or its core wire is open cir cuit it can be a cause of electric shock hazards Never use imper...

Страница 49: ...s cutoff and FAIL alarm is gen erated If no FAIL judgment is made after preset period has elapsed PASS signal is generated High limit reference value CUTOFF CURRENT 0 5 1 2 4 8 10 100 mA 7 values By c...

Страница 50: ...asuring current 1 0 mA Short circuit current 12 mA or less Effective Measuring Ranges 500 V range 1 M to 1 000 M 1 000 V range 2 M to 2 000 M Values center of scale 500 V range 20 M 1 000 V range 50 M...

Страница 51: ...and FAIL alarm Signal Name Conditions for Signal Generation Type of Signals TEST ON signal Delivered during entire test on period Make contact signal and lamp PASS signal Delivered when PASS judgment...

Страница 52: ...Conforms to the requirements of the following directive and standard EMC Directive 89 336 EEC EN61326 EN61000 3 2 EN61000 3 3 Under following conditions 1 Used HV test leadwire TL01 TOS 2 No discharge...

Страница 53: ...TOS8870A SPECIFICATIONS 41 7 4 Outline Drawing Unit mm Fig 7 1...

Страница 54: ...time with your both hands thereby enhancing the operating safety Functions of switches OPERATE switch This switch enables when ON or disables when OFF the START switch or switches START switch The te...

Страница 55: ...if you do this the DUT may be damaged If you disconnect the probe from the DUT while letting it deliver the test voltage the items of the tested circuit may remain charged up to the high test voltage...

Страница 56: ...TOS8870A OPERATION MANUAL...