TOS8030
Status Signal Output 43
Chapter 7
Status Signal Output
Describes the status signal outputs (SIGNAL OUT).
7.1
Five Signal Outputs
The following status signals are output from the SIGNAL OUT terminals on the
rear panel. These signals are output in sync with the states defined in “5.2.1 Five
States.” For more information on each state, see “5.2.1 Five States” (page 32).
7.2
Using the SIGNAL OUT Terminals
7.2.1
Description of the Terminals
The SIGNAL OUT terminals are circuits in which a contact is closed when a signal
is output (make contact signal), and there is no power supply in the internal circuits
of SIGNAL OUT terminals. The terminals cannot drive a load without a power sup-
ply. See the internal circuit for each pair of terminals shown in Fig. 7-1.
Fig. 7-1
Connecting a Wire to the SIGNAL OUT Terminals
READY signal
This signal is output when the tester is in a READY state.
H.V ON signal
This signal is output during testing or when high voltage remains in the output terminals
(the DANGER lamp remains lit).
PASS signal
This signal is output when the test result is determined to be PASS. A PASS signal is
output for approx. 200 ms, after which the tester shifts to a READY state, outputting a
READY signal.
FAIL signal
This signal is output when the test result is determined to be FAIL. A FAIL signal is
output continuously until the tester enters a READY state or PROTECTION state.
PROTECTION
signal
This signal is output when the tester is in a PROTECTION state.
Wiring material
AWG20 to 26 (1.5 mm
2
)
Insert a wire into the
terminals, pressing on
this part using a
screwdriver or the like.
Internal circuit of
each pair of terminals
Rating of make
contact signal:
1 A/30 Vdc
Turn the POWER switch
off before connecting
a wire to the SIGNAL
OUT terminals.