32 Panel Operations
TOS8030
5.2.1
Five States
This tester recognizes five tester states: TEST, PASS, FAIL, PROTECTION, and
READY, which are defined below:
■
TEST state
State from start of to suspension or completion of testing
The TEST and DANGER lamps remain lit to indicate that a test voltage is being
output. An H.V ON signal is generated.
■
PASS state
Indicates a state in which testing has ended and the result has been determined to be
PASS.
The PASS lamp lights, a buzzer sounds, and a PASS signal is generated to indicate a
PASS state for approx. 200 ms. The tester then transitions to a READY state.
■
FAIL state
Indicates a state in which testing has ended and the result has been determined to be
FAIL.
The FAIL lamp lights, a buzzer sounds, and a FAIL signal is generated to indicate a
FAIL judgment. The FAIL state can be canceled by pressing the STOP switch, after
which the tester transitions to a READY state.
■
PROTECTION state
Indicates a state in which the protection function is activated.
The PROTECTION lamp lights, and a PROTECTION signal is generated to indi-
cate that the tester is in the PROTECTION state. A code representing the cause of
protection function activation is also displayed on the voltmeter in this state. For
more information, see “5.2.2 Events or Conditions That Can Activate the Protection
Function” (page 33).
The PROTECTION state can be canceled by pressing the STOP switch (or by input-
ting a STOP signal), after which the tester transitions to a READY state.
■
READY state
A READY signal is generated in this state. All four lamps (TEST, PASS, FAIL, and
PROTECTION) remain unlit.
Press the START switch in this state to start testing.