PWR Series Setup Guide
15
Load Wire
Limitations of the sensing function
All wires have resistance. The voltage drop in wires becomes
greater as the wire becomes longer or the current becomes larger.
This results in the voltage applied at the load end to be smaller.
The PWR has a sensing function that compensates for this voltage
drop up to approximately 0.6 V for a single line. If the voltage drop
exceeds this level, wires having a greater sectional area should be
used.
Nominal cross-sectional area of wires and allowable currents
Nominal Cross-
Sectional Area
[mm
2
]
AWG
(Reference
Cross-Sectional
Area) [mm
2
]
Allowable Cur-
rent* [A]
(Ta = 30 ºC)
Current Recom-
mended by
Kikusui [A]
2
14
(2.08)
27
10
3.5
12
(3.31)
37
-
5.5
10
(5.26)
49
20
8
8
(8.37)
61
30
14
6
(13.3)
88
50
22
4
(21.15)
115
80
30
2
(33.62)
139
-
38
1
(42.41)
162
100
50
1/0
(53.49)
190
-
60
2/0
(67.43)
217
-
80
3/0
(85.01)
257
200
100
4/0
(107.2)
298
-
125
-
-
344
-
150
-
-
395
300
200
-
-
469
-
*Excerpts from Japanese laws related to electrical equipment.