PLZ-4W
Installation and Preparation 2-17
2.8.3
Connection to the Load Input Terminal on the Front Panel
The load input terminal on the front panel can be used to easily connect the equip-
ment under test and the PLZ-4W.
• Do not touch the load input terminal while the PLZ-4W is turned ON, as it
may lead to electric shock.
• The load input terminals on the front and rear panels are coupled inside
the PLZ-4W. A voltage applied to one end will appear on the other end.
• Do not connect equipment to the load input terminals on the front and rear
panels simultaneously, as it cause damage.
• There is a danger of breakdown. Do not connect the equipment under test
to the load input terminal when the load is turned on.
• There is a danger of overheat. Attach crimping terminal to the wire and
use the set of screws that came with the package for connection.
• The speci
fi
cations of the PLZ-4W are for the load input terminal on the rear panel.
• The maximum input current to the load input terminal on the front panel is 66 A.
It is automatically limited to 66 A.
1.
Turn off the POWER switch.
2.
Check that the output of the equipment under test is off.
3.
Connect the load wire to the load input terminal on the front panel.
For the connection procedure of the load wire, see Fig. 2-17.
4.
Connect the load wire to the output terminal of the equipment under test.
5.
Check the polarity of the connection.
Fig. 2-17
Connection to the load input terminal on the front panel
WARNING
CAUTION
NOTE
Spring washer (M6)
Crimp terminal
Cover
Knob
M6
Attach so that the wire
passes through this section.
Содержание PLZ-4W Series
Страница 10: ...VIII Contents PLZ 4W ...
Страница 20: ...1 10 General Information PLZ 4W ...
Страница 38: ...2 18 Installation and Preparation PLZ 4W ...
Страница 90: ...5 30 Basic Operation PLZ 4W ...
Страница 192: ...9 10 Specifications PLZ 4W ...
Страница 214: ...A 22 Appendix PLZ 4W ...