PCZ1000A
3-11
3
Ope
ration
3.5.1
CC Mode Operation
In constant current (CC) mode, set a current limit [A] value.
Current limit setting range: 0.00 Arms to 10 Arms
Fig.3-3 shows the front panel display when in constant current mode. The input
terminal measurements are displayed on the upper side.
Fig.3-3
Display in CC Mode
1
If the LED above the LOAD key is on, press the LOAD key to turn it off.
2
Press the I SET key to enable CC mode.
The LED above the I SET key illuminates.
3
Turn the rotary knob quickly clockwise to set the current limit value
approximately.
4
Turn the rotary knob slowly clockwise to set the exact current limit
value.
5
Press the LOAD key to change to the load-on state.
The LED above the LOAD key illuminates and current flow starts.
0.00Ar 0.0Ap 100.1V
-- ISET 0.00A --
RMS current value
Peak current value
RMS voltage value
Current limit
0.00Ar 0.0Ap 100.1V
-- ISET 0.00A --
0.00Ar 0.0Ap 100.1V
-- ISET 4.50A --
0.00Ar 0.0Ap 100.1V
-- ISET 5.00A --
5.00Ar 7.1Ap 100.0V
-- ISET 5.00A --
Содержание PCZ1000A
Страница 4: ...ii PCZ1000A This page is intentionally blank ...
Страница 14: ...xii PCZ1000A Front panel 18 17 Control panel Control panel 1 Display Display 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Страница 35: ...Operation This chapter describes each mode and function of the operation ...
Страница 82: ...5 4 PCZ1000A This page is intentionally blank ...
Страница 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Страница 88: ...6 6 PCZ1000A This page is intentionally blank ...
Страница 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Страница 99: ......