Maintenance
Periodic inspections and maintenance are required to maintain the product’s original
performance over the long term.
•
Be sure to disconnect the test probe from the Tester before starting inspec-
tions or maintenance.
•
Avoid volatile solvents such as thinner and benzene, which may discolor the
probe’s surface coating, erase printed characters, etc.
Cleaning
If the test probe becomes soiled, moisten a piece of soft cloth with a mild, dilute detergent
and gently wipe the probe.
Periodic Inspection
The cable of the test probe, contact pin, and LOW test lead are consumables. We
recommend an annual inspection by one of service engineers. To have an inspection
performed, contact your Kikusui distributor or agent.
Specifications
*1:The rated maximum voltage/current refers to the maximum voltage/
current the test probe is capable of handling within the given operating
temperature and humidity ranges.
LP01-TOS
LP02-TOS
Rated maximum voltage (*1)
30 V
Rated maximum current (*1)
30 Aac
60 Aac
Residual resistance
10 m
or less
Operating environment
Indoor use, Altitude: up to 2000 m
Operating temperature and
relative humidity range
0
C to 40
C
20 % to 80 % (no condensation)
Storage temperature and
relative humidity range
-20
C to 70
C
90 % or less (no condensation)
Insulation
resistance
Between contact
pin and signal cable
30 M
or more (500 Vdc)
Withstanding
voltage
Between contact
pin and signal cable
500 Vac for 5 seconds, 2 mA or less
Cable length
Approx. 2 m
Physical dimensions
28 (W) x 45.5 (H) x 226 (D) mm
Weight
Approx. 500 g
Approx. 700 g
Accessories
LOW test lead: 1 (1.5 m)
Operation manual: 1 copy
Description