Maintenance

Periodic  inspections  and  maintenance  are  required  to  maintain  the  product’s  original

performance over the long term.

Be sure to disconnect the test probe from the Tester before starting inspec-
tions or maintenance.

Avoid volatile solvents such as thinner and benzene, which may discolor the

probe’s surface coating, erase printed characters, etc.

Cleaning

If the test probe becomes soiled, moisten a piece of soft cloth with a mild, dilute detergent

and gently wipe the probe.

Periodic Inspection

The  cable  of  the  test  probe,  contact  pin,  and  LOW  test  lead  are  consumables.  We

recommend  an  annual  inspection  by  one  of  service  engineers. To  have  an  inspection

performed, contact your Kikusui distributor or agent.

Specifications

*1:The  rated  maximum  voltage/current  refers  to  the  maximum  voltage/

current the test probe is capable of handling within the given operating

temperature and humidity ranges.

LP01-TOS

LP02-TOS

Rated maximum voltage (*1)

30 V

Rated maximum current (*1)

30 Aac

60 Aac

Residual resistance

10 m

 or less

Operating environment

Indoor use, Altitude: up to 2000 m

Operating temperature and 

relative humidity range

C to 40



20 % to 80 % (no condensation)

Storage temperature and 

relative humidity range

-20 

C to 70 

90 % or less  (no condensation)

Insulation 

resistance

Between contact 

pin and signal cable

30 M

 or more (500 Vdc)

Withstanding 

voltage

Between contact 

pin and signal cable

500 Vac for 5 seconds, 2 mA or less

Cable length

Approx. 2 m

Physical dimensions

28 (W) x 45.5 (H) x 226 (D) mm

Weight

Approx. 500 g

Approx. 700 g

Accessories

LOW test lead: 1 (1.5 m)

Operation manual: 1 copy

Description

Содержание LP01-TOS

Страница 1: ...r earth continuity testers The test probe is designed specifically for use with a KIKUSUI Earth Continu ity Tester It cannot be used with other earth continuity testers even if the connector appears t...

Страница 2: ...from the switch will stop testing Contact Pin Test current is delivered to this pin contacting a test point of the DUT Holder This part holds the contact pin OUTPUT LED This red LED indicates that te...

Страница 3: ...epairs Test probe cable and LOW test lead Check for broken sheathing and loose or cracked connectors Confirm electrical continuity Start switch Confirm that the switch operates smoothly when pressed C...

Страница 4: ...OUTPUT LED does not light The on off status of the OUTPUT LED of the test probe is inconsistent with that of the TEST LED on the Tester The OUTPUT LED remains lit even when you release your finger fr...

Страница 5: ...e LOW terminal of the Tester 4 Connect the each cable of the test probe to the HIGH and REMOTE termi nal of the Tester Connecting the remote control cable starts or stops testing as indicated below Te...

Страница 6: ...nect the alligator clip of the LOW test lead to the protective conductor terminal of the DUT 2 Apply the contact pin of the test probe to the test point of the DUT 3 Depress the START switch This caus...

Страница 7: ...i distributor or agent Specifications 1 The rated maximum voltage current refers to the maximum voltage current the test probe is capable of handling within the given operating temperature and humidit...

Страница 8: ...e equipment and quantity SJ T 11364 GB T 26572 GB T 26572 Environment friendly Use Period Pb Hg Cd Cr VI PBB PBDE LP01 TOS LP02 TOS KIKUSUI ELECTRONICS CORP www kikusui co jp en 1 1 3 Higashiyamata Ts...

Отзывы: