LP01/02-TOS

5

Connecting to the Earth Continuity Tester

Improper terminal connections can result in inaccurate measure-
ments and burns or injury resulting from heat generated by con-
tact resistance at the terminals.

1.

Turn off the POWER switch of the Tester.

2.

Remove the short-circuit bars connecting the OUTPUT and SAMPLING ter-
minals.

3.

Connect the LOW test lead to the LOW terminal of the Tester.

4.

Connect the each cable of the test probe to the HIGH and REMOTE termi-
nal of the Tester.

Connecting the remote control cable starts or stops testing as indicated below:

Tester’s Panel Operation

Operation of Test Probe

Starting test

Disabled

Depress the START switch.

Stopping test

Press the STOP switch.

Release your finger from the 

START switch.

WARNING

To SAMPLING
HIGH terminal

Figure shows connections of the 
TOS6210 with the LP02-TOS.

To the REMOTE terminal

To OUTPUT
HIGH terminal

To SAMPLING
LOW terminal

To OUTPUT
LOW terminal

TOS6210

Содержание LP01-TOS

Страница 1: ...r earth continuity testers The test probe is designed specifically for use with a KIKUSUI Earth Continu ity Tester It cannot be used with other earth continuity testers even if the connector appears t...

Страница 2: ...from the switch will stop testing Contact Pin Test current is delivered to this pin contacting a test point of the DUT Holder This part holds the contact pin OUTPUT LED This red LED indicates that te...

Страница 3: ...epairs Test probe cable and LOW test lead Check for broken sheathing and loose or cracked connectors Confirm electrical continuity Start switch Confirm that the switch operates smoothly when pressed C...

Страница 4: ...OUTPUT LED does not light The on off status of the OUTPUT LED of the test probe is inconsistent with that of the TEST LED on the Tester The OUTPUT LED remains lit even when you release your finger fr...

Страница 5: ...e LOW terminal of the Tester 4 Connect the each cable of the test probe to the HIGH and REMOTE termi nal of the Tester Connecting the remote control cable starts or stops testing as indicated below Te...

Страница 6: ...nect the alligator clip of the LOW test lead to the protective conductor terminal of the DUT 2 Apply the contact pin of the test probe to the test point of the DUT 3 Depress the START switch This caus...

Страница 7: ...i distributor or agent Specifications 1 The rated maximum voltage current refers to the maximum voltage current the test probe is capable of handling within the given operating temperature and humidit...

Страница 8: ...e equipment and quantity SJ T 11364 GB T 26572 GB T 26572 Environment friendly Use Period Pb Hg Cd Cr VI PBB PBDE LP01 TOS LP02 TOS KIKUSUI ELECTRONICS CORP www kikusui co jp en 1 1 3 Higashiyamata Ts...

Отзывы: