Connecting to the Earth Continuity Tester
•
Improper terminal connections can result in inaccurate measure-
ments and burns or injury resulting from heat generated by con-
tact resistance at the terminals.
1.
Turn off the POWER switch of the Tester.
2.
Remove the short-circuit bars connecting the OUTPUT and SAMPLING ter-
minals.
3.
Connect the LOW test lead to the LOW terminal of the Tester.
4.
Connect the each cable of the test probe to the HIGH and REMOTE termi-
nal of the Tester.
Connecting the remote control cable starts or stops testing as indicated below:
Tester’s Panel Operation
Operation of Test Probe
Starting test
Disabled
Depress the START switch.
Stopping test
Press the STOP switch.
Release your finger from the
START switch.
WARNING
To SAMPLING
HIGH terminal
Figure shows connections of the
TOS6210 with the LP02-TOS.
To the REMOTE terminal
To OUTPUT
HIGH terminal
To SAMPLING
LOW terminal
To OUTPUT
LOW terminal
TOS6210