128
KHA3000
Starting the Test by Pressing the START Key
(61000-3-2)
1
Press the VIEW key.
The HA-Observation and Analysis display appears.
2
Select a view type for the test.
To observe the harmonic current of each order, select 2D Harmonics. Some displays
can be selected only after the test ends.
3
Press the START key.
The dialog box for confirming the line impedance is displayed.
If the optional current sensor is not connected or recognized although "Sensor" is selected
in the Current Input Terminal setting, a confirmation dialog will be presented.If the sum (
) of
the real power of each input phase is negative, a confirmation dialog will be presented.
4
After checking presence/absence of impedance and the setting of line
impedance, select the F1 key (OK).
If the F2 key (Cancel) is selected, the test start is canceled. To start the test, return to
Step 3 and press the START key.
The test is started.
Aborting the Test by Pressing the STOP Key
�
Press the STOP key during the test.
The test ending menu (p. 129) is displayed. Test results are aborted. The subsequent
operations are the same as for “
When the Test Time Expires (61000-3-2)
.”
When the Test is Aborted Automatically
The test is aborted automatically when a measured value exceeds the voltage or current
range, synchronization of PLL fails, or overheating of the current detecting portion is
detected.
The test ending menu (p. 129) is displayed. Test results are aborted. The subsequent
operations are the same as for
When the Test Time Expires (61000-3-2)
.”
See
If JIS 61000-3-2 2005 or JIS 61000-3-2 2019 is selected and the reference impedance is
used, check the AC power source and record before the test. If the START key is pressed
without doing this, an AC power source check dialog will be displayed.
See
See
Display during test
Test status display Switches from
“
Set
”
to
“
Test.
”
The progress bar extends from left to right. When it reaches the right end,
the test ends.
Time display
Remaining test time
View type
The view type that is set in the HA-Observation and Analysis display (HA-VIEW)
Display after the test ends
Test status display Switches from
“
Test
”
to
“
Analysis.
”
The progress bar extends to the right end.
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......