DSI1020/ DSI3020
21
Installing the Test System
1
Check that the POWER switches of all devices are turned off ( ), and
shut off the power supply from the switchboard.
2
Check that the
terminal (protective conductor terminal) of each AC
power supply’s INPUT terminal block is grounded.
3
Unfasten the screws from the terminal cover attached to the lower part
of the rear panel of this product.
4
Connect the OUTPUT terminal blocks of the AC power supplies to the
INPUT terminal block of this product using wires.
“DSI1020 test circuits: two types” (p. 22)
and
, and select the test circuit appropriate for your EUT.
For connection details, see the AC power supply manual.
5
Connect the OUTPUT terminal block of this product to the INPUT termi-
nal block of the EUT using wires.
If necessary, attach relay terminals between this product and the EUT.
6
Attach the terminal cover that you removed in step 3.