Getting Started and Troubleshooting Guide

Содержание x-series

Страница 1: ...This documentation provides information for the following X Series Instruments UXA Signal Analyzer N9041B Getting Started and Troubleshooting Guide Keysight X Series Signal Analyzer ...

Страница 2: ...are documentation No additional government requirements beyond thoseset forth in the EULAshall apply except to theextent that thoseterms rights or licenses areexplicitly required from all providers of commercial computer softwarepursuant to theFAR and theDFARS and areset forth specifically in writing elsewherein theEULA Keysight shall beunder no obligation to update reviseor otherwisemodify theSof...

Страница 3: ...formation 16 Protecting against electrostatic discharge 17 Test equipment and ESD 17 Additional information about ESD 17 2 Front and Rear Panel Features 19 Front Panel Features 20 Display Features 23 Menu Panel 26 Rear Panel Features 28 Front and Rear Panel Symbols 30 3 Instrument Operating System 31 Microsoft Windows 32 Windows Help and Support Center 32 Installed Software 33 Signal analyzer soft...

Страница 4: ...s Protection 46 Spyware Protection 47 System Maintenance 48 Backup 48 System Restore 48 Disk Defragmenting 48 USB Connections 49 Disk Drive Partitioning and Use 50 Backing up the factory calibration data 51 Disk Drive Recovery Process 52 Updating the software 54 4 Using Window Tools 57 Navigating Windows Without a Mouse 58 Remote Desktop Using the X Series Signal Analyzers Remotely 59 Overview of ...

Страница 5: ... 73 5 Troubleshooting 75 Check the Basics 76 Problems with Microsoft Windows 7 78 Returning an Analyzer for Service 79 Calling Keysight Technologies 79 Locations for Keysight Technologies 80 Read the Warranty 80 Service Options 81 Packaging the Instrument 82 6 Keysight Connector Kit 83 Connector Kit Overview 84 Use of the Test Port Adaptors on Instrument RF Input 2 85 Input Connector Vise Assembly...

Страница 6: ...6 Getting Started and Troubleshooting Guide ...

Страница 7: ...e found in this section Initial inspection on page 8 Instrument Location and Rack Mounting Requirements on page 9 Turning on the Analyzer the First Time on page 10 Anti Virus Software and Firewalls on page 13 Instrument Information on page 14 Keysight X Series Signal Analyzer Getting Started and Troubleshooting Guide 7 ...

Страница 8: ... Contents Verify the shipping container contents using the box contents list Shipping Problems If the shipping materials are damaged or the contents of the container are incomplete Contact the nearest Keysight Technologies office Keep the shipping materials for the carrier s inspection If you must return an analyzer to Keysight Technologies use the original or comparable shipping materials See Ret...

Страница 9: ...ke sure the exhaust air from the first instrument is directed away from the inlet of the second unit If the pre heated air from the first instrument is directed into the second instrument it can cause excessive operating temperatures in the second unit and can cause instrument failures The analyzer draws air in from the left side and exhausts air from the right side When installing the product in ...

Страница 10: ...located in the lower left corner of the analyzer s front panel to turn the analyzer on The analyzer can require more than 5 minutes to power on The Keysight Technologies screen appears followed by a screen that allows you to select Windows 7 or the Recovery option 2 Launch Microsoft Setup The analyzer performs the following steps Windows 7 Startup window Black screen Windows Please wait message wi...

Страница 11: ...h time the analyzer is turned on No application will start while this message is displayed Before continuing make sure that you carefully read the Anti Virus message and determine what action is appropriate 4 Disable the Anti Virus message Select the check box and click Continue Messages similar to the following continue to appear Several required processes continue then the application initialize...

Страница 12: ...tem b Verify that the purchased application s appear in the list If you require further assistance contact the Keysight support team Online assistance http www key sight com find assist 8 View a signal a Press Input Output RF Calibrator 50 MHz b Press Freq Center Freq 50 MHz c Press Span 50 MHz This routes the internal 50 MHz signal to the analyzer input The 50 MHz reference signal appears on the ...

Страница 13: ... that you install anti virus software if your analyzer is connected to the LAN Check with your IT department to see what they recommend The analyzer is shipped with the Windows 7 firewall enabled Do not modify the default network settings as this may cause problems with the operating system of the analyzer Getting Started and Troubleshooting Guide 13 ...

Страница 14: ... turning on the analyzer you must connect its protective earth terminals to the protective conductor of the main power cable Insert the main power cable plug into a socket outlet that has a protective earth contact only DO NOT defeat the earth grounding protection by using an extension cable power cable or autotransformer without a protective ground conductor The Mains wiring and connectors shall ...

Страница 15: ...input level of 30 dBm average total power or 0 2 Vdc The Input 2 circuitry of the analyzer can be damaged by applying signals that exceed the maximum safe input level of 5 dBm average total power or 0 Vdc Refer to the analyzer s specification guide for more details regarding the Maximum Safe Input Level Repairing damage to the input circuitry can be expensive If the analyzer will be used to measur...

Страница 16: ...rmful if absorbed through the skin Wash thoroughly after handling In case of spill soak up with sand or earth Flush spill area with water Dispose of isopropyl alcohol in accordance with all applicable federal state and local environmental regulations Battery Information The analyzer uses a lithium battery located on the CPU board This is not an operator replaceable part See Returning an Analyzer f...

Страница 17: ...st strap before touching the center pin of any connector and before removing any assembly from the analyzer Be sure that all instruments are properly earth grounded to prevent build up of static charge Perform work on all components or assemblies at a static safe workstation Keep static generating materials at least one meter away from all components Store or transport components in static shieldi...

Страница 18: ...1 Quick Start Instrument Information 18 Getting Started and Troubleshooting Guide ...

Страница 19: ...ures This section describes the following features Front Panel Features on page 20 Display Features on page 23 Front and Rear Panel Symbols on page 30 Keysight X Series Signal Analyzer Getting Started and Troubleshooting Guide 19 ...

Страница 20: ...well as the Input Output variables Does not affect System variables 3 Save Recall Keys Save enables you to save states traces screen images and other items from the analyzer to files on the analyzer s internal storage to removable devices and to directories on the network Recall enables you to recall previously saved states traces and other items to the analyzer from files on the analyzer s intern...

Страница 21: ...ailable in the Utility section Numeric key pad System Help Local Cancel Esc Back space Delete Control Alt Undo Redo Touch On Off Onscreen Keyboard Tab 9 Input 1 2 4 mm m Connector for inputting external signals 3 Hz to 50 GHz Make sure that the total power of all signals at this analyzer input does not exceed 30 dBm 1 watt 10 USB Connectors Standard USB 2 0 ports Type A Connect to external periphe...

Страница 22: ...W with 0 dB attenuation 15 Power Standby On Turns the analyzer on A green light indicates power on A yellow light indicates standby mode The front panel switch is a standby switch not a LINE switch disconnecting device The analyzer continues to draw power even when the front panel switch is in standby The main power cord can be used as the system disconnecting device It disconnects the mains circu...

Страница 23: ...e display are tabs one for each measurement screen you have defined Tap the sign to clone the current measurement which can be changed once it is created Tap the current screen tab or press the Mode Meas hardkey to display the following dialog This dialog allows you to choose a Mode Measurement and View Getting Started and Troubleshooting Guide 23 ...

Страница 24: ...bar they scroll left and right using the arrows to the left and right of the tabs You switch screens by touching the tabs To view multiple screens press the icon on the Bottom Bar Meas Bar The Meas Bar shows general measurement settings and information The annotations on this bar can be used to change settings Tap anywhere in the annotation box to access the drop down panel that contains relevant ...

Страница 25: ... the reference level Markers may be moved by dragging them to the desired location Touch and hold simulates a right click Swipe scroll allows you to view information that extends beyond the window area When you begin to scroll the scroll bar appears and fades once you stop scrolling The annotation drop downs in the window area allow you to change parameters The window title drop down allows you to...

Страница 26: ...ey drop down panel contains the measurement controls These are the same as the hardkeys in the shaded area of the keypad You can select functions using either the keypad or the drop down menu Notice that the Frequency panel has a Settings tab on the right side Other panels may have multiple tabs The tabs access controls for the particular parameter noted on the tab 26 Getting Started and Troublesh...

Страница 27: ...2 Front and Rear Panel Features Display Features Bottom Bar The bottom bar contains several icons that access various controls Getting Started and Troubleshooting Guide 27 ...

Страница 28: ...An output of the analyzer internal 10 MHz frequency reference signal It is used to lock the frequency reference of other test equipment to the analyzer 5 SNS Series Noise Source For use with Keysight N4000A N4001A N4002A Smart Noise Sources SNS 6 Noise Source Drive 28 V Pulsed For use with Keysight 346A 346B and 346C Noise Sources 7 TRIGGER 1 IN Allows external triggering of measurements 8 TRIGGER...

Страница 29: ...ers 17 VGA Allows connection of an external VGA monitor 18 DisplayPort Used for video output Accepts a standard mini display port connector or adaptor for connection of an external monitor 19 LAN USB A TCP IP Interface that is used for remote analyzer operation USB 2 0 port Type A Connect to external peripherals such as a mouse keyboard printer DVD drive or hard drive 20 USB Connector USB 3 0 port...

Страница 30: ...B 001 This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme a la norme NMB du Canada The CSA mark is a registered trademark of the CSA International This symbol indicates separate collection for electrical and electronic equipment mandated under EU law as of August 13 2005 All electric and electronic equipment are required to be separated from normal waste for disposal Refe...

Страница 31: ...USB mouse and external keyboard For more useful shortcuts see Navigating Windows Without a Mouse on page 58 The following topics can be found in this chapter Microsoft Windows on page 32 Installed Software on page 33 Customer Installation of Software on page 34 User Accounts on page 35 Keysight X Series Analyzer Licensing Options on page 37 Licensing New Measurement Application Software on page 38...

Страница 32: ...dows Help and Support Center Throughout this chapter you will find references to the Windows Help and Support Center To access the Windows Help and Support Center follow the instructions below either on your PC or on the instrument itself You must have an Internet connection to access the Windows Help and Support Center If the xSA is on screen tap the icon in the lower left corner of the display T...

Страница 33: ...are installed by the factory at the time of manufacture even if the licenses have not been purchased You may purchase additional licenses at a later date Vector signal analyzer software The 89600 VSA software is installed in the signal analyzer This software was installed by the factory at the time of manufacture even if the license was not purchased If the license was not purchased VSA will run i...

Страница 34: ...mance of the analyzers with non approved software installed Before installing any additional programs on the instrument you should exit the Signal Analyzer application Also you must not remove any applications or programs that were installed on the instrument when it was shipped from the factory If you install programs other than those that Keysight has tested it could cause problems with the inst...

Страница 35: ...sing Device Manager Change the time and date Run any application Instrument accounts The default user account that ships from the factory is Instrument with the password measure4u This user is a member of the standard Users group Using the Instrument account you may perform the following operations Configure network and printer access although not local printer access Access files on the instrumen...

Страница 36: ...security granted to any new user accounts created For example the level of security can be assigned as administrator power user standard user backup operators User names are not case sensitive but passwords are case sensitive It is Keysight s expectation that each user s My Documents folder is mapped to the D drive This is to avoid overwriting the user s data in the event the Instrument Recovery m...

Страница 37: ...d to a specific instrument model and serial number Transportable Perpetual licenses are identified in the product structure by a T in the second character and P in the third character of the option designator Example N9068C 2TP Transportable Perpetual licenses require a connection to the Keysight server only for managing the check in out of the license The Keysight licensing server also provides f...

Страница 38: ...ation installation Installation procedure over USB Step Action Notes 1 Redeem the Option Upgrade Entitlement Certificate Follow the instructions on the Certificate After redeeming your Option Upgrade Entitlement Certificate you will receive an e mail with an attached License File 2 Save the license file Save the lic file to the root directory of a USB storage device 3 Load the license file Connect...

Страница 39: ... If you require further assistance please contact the Keysight support team Online assistance http www keysight com find assist If you do not have access to the Internet contact your local Keysight Technologies Sales and Service Office or if in the United States call 1 800 829 4444 Getting Started and Troubleshooting Guide 39 ...

Страница 40: ...ion in case an Instrument Recovery is performed and the configuration is reset Some of these actions can only be performed with Administrator privileges You May Use This Feature To Do This Configure Microsoft Windows Automatic Updates Microsoft recommends that you always get the latest critical Windows updates to ensure that the instrument s Windows operating system is protected If the instrument ...

Страница 41: ...ngs that must not be changed Avoid changing the settings described below available from the Windows Control Panel Changes to these settings may degrade instrument performance screen displays and measurement speed Do NOT Use This Feature To Do This Do not change Power Options If you click on Advanced System Settings a dialog will open called System Properties On this dialog there is a tab called Ha...

Страница 42: ...s for the NET Framework The Microsoft Visual J NET Redistributable Package 1 1 Programs that begin with Keysight The Adobe Acrobat reader Modify The I O Library GPIB27 GPIB28 interfaces shown as configured Instrument I O in the Connection Expert or I O Config Autoplay Autorun Since the introduction of Windows XP the term Autoplay sometimes also called Autorun has come to be associated with the fea...

Страница 43: ...gating Windows Without a Mouse on page 58 When setting up a new printer you may need to load the printer driver unless you are using a network printer that your IT department has set up to include the driver The manufacturer of the printer supplies the driver software and process That may require that you attach an external USB disk drive An alternative is to connect the instrument to the LAN and ...

Страница 44: ...pport Center IP Address Gateway The instrument is preconfigured to obtain an IP Address via DHCP You can change the IP Address and Gateway as you desire The Windows Help and Support Center describes how to configure the LAN Connecting to a Network Shared Folder The instrument contains standard Windows networking The time required to authenticate is dependent on your LAN infrastructure You may have...

Страница 45: ...ng LAN When the Map Network Drive window appears browse to the correct folder select both check boxes and click Finish In Windows 7 there is no visual indication that authentication is in progress Getting Started and Troubleshooting Guide 45 ...

Страница 46: ... has internet access the instrument default is set to automatically check for critical Windows Updates and notify you You can change the configuration of the Microsoft Automatic Updates You can choose not to have automatic updates However if you do this then you should manually update Windows periodically by accessing Internet Explorer and selecting Windows Update from the Tools menu Be aware that...

Страница 47: ... no anti spyware software installed on the instrument This should not be a problem if you do not use the instrument for a lot of internet browsing Having spyware in the instrument could have an impact on the instrument performance Getting Started and Troubleshooting Guide 47 ...

Страница 48: ...ftware See Customer Installation of Software on page 34 for more information When performing backups we recommend that you backup the data to an external storage device connected to the network or one of the instrument s USB connectors Also you should perform backups at times when the instrument is not being used for normal operations as it may impact the instrument s overall performance System Re...

Страница 49: ...ries B port and is used for controlling the instrument over USB Information to help you program your instrument is documented in the X Series Programmer s Guide The instrument USB device driver included in the instrument software supports the test and measurement industry standard USBTMC USB488 device class In addition the IO Libraries CD that was included with your instrument contains USB Host dr...

Страница 50: ...ument Recovery process You must reload any other software that you have added in the instrument The D partition is reserved for data storage The User Accounts that are configured by Keysight have their My Documents folder mapped to the D drive This is for the convenience of backing up the measurement data You should always back up the data on the D drive to an external device This allows you to re...

Страница 51: ...information window appears that informs you that the instrument must shut down to complete the backup Press OK to force shut down and proceed Press Cancel to exit 2 Select OK The Alignment Data Wizard will appear 3 Follow the on screen instructions to backup the calibration data to the USB drive and save the file The default location is an internal flash SD card but you can choose another location...

Страница 52: ...d to be re installed System configurations for example user accounts windows configurations that have been made after the instrument was shipped from the factory Thus following an Instrument Recovery configuration changes will have to be performed The Instrument Recovery overwrites the contents of the C partition It does not affect the D or E partitions It is recommended that the customer use a re...

Страница 53: ...nstructions to recover the image of the C drive Press 2 then press Enter to select the recovery 4 A warning message appears Press Enter to start the recovery which may take up to 25 minutes to complete 5 Press Enter to exit and reboot the instrument once this portion of the recovery has completed Getting Started and Troubleshooting Guide 53 ...

Страница 54: ... code resident in the instrument s various programmable gate arrays to be in sync with the restored system software Step Notes 1 Log out of the default user instrument Select Start select Log Off from the Shut down drop down menu 2 At the log in prompt enter User Name administrator Password Keysight4u 3 Navigate to C Drive Click Start My Computer C Drive and open the Temp folder 4 Locate the softw...

Страница 55: ...s damage may occur If any pop up windows appear click OK or Ignore to proceed 8 When the installation has finished select Yes I want to restart my computer now Finish 9 After the instrument restarts the newly installed version of the X Series instrument software will run It is possible a newer version of the instrument software is available on the web at http www keysight com find xseries_software...

Страница 56: ...3 Instrument Operating System Disk Drive Recovery Process 56 Getting Started and Troubleshooting Guide ...

Страница 57: ...mation Your version of Windows may not match these instructions exactly You need an external keyboard and mouse to fully use these features Navigating Windows Without a Mouse on page 58 Remote Desktop Using the X Series Signal Analyzers Remotely on page 59 Embedded Web Server Using the X Series Signal Analyzers Remotely on page 66 Windows Shortcuts and Miscellaneous Tasks on page 72 Keysight X Ser...

Страница 58: ...on up in the menu In dialog box selects an option button Down Arrow In pull down menu Moves to next selection down in the menu In dialog box selects an option button Tab In dialog box moves to the next previous field Del Deletes the currently selected item Alt Tab Switches between the next previous Application Alt Enter Shows the Properties of the currently selected item Alt Esc Cycles through ite...

Страница 59: ...Windows 7 help documentation for more information As Windows evolves these instructions may no longer be exact You need an external keyboard and mouse to fully use this functionality Overview of Remote Desktop operation Using the Remote Desktop functionality of the instrument allows you to control and interact with the instrument from a remote computer as though you were sitting in front of the in...

Страница 60: ...ctivity Client software so no additional setup is required Remote computer running another version of Windows You can use any version of Windows to install and run the Client software for Remote Desktop Connectivity However you need to have available a Windows installation CD ROM because that contains the Client software The following instructions relate to software provided by Microsoft Corporati...

Страница 61: ...cating the name from the Windows desktop with a mouse Running a Remote Desktop session Initializing a Remote Desktop session To initialize a Remote Desktop Session you need to know the Computer Name of the instrument After setting up the remote computer for Remote Desktop Connectivity as described in Remote Desktop Using the X Series Signal Analyzers Remotely on page 59 you are ready to start a Re...

Страница 62: ...splay appears on the screen of the remote computer A mouse connected to the remote computer is used to navigate the user interface and change instrument settings Windows Remote Desktop Options Step Notes 1 On the Remote Desktop Connection menu click Options The Options dialog has several tabs Generally the default settings are correct 2 Under the General tab ensure that the Computer name User name...

Страница 63: ...portions of the display are not accessible Under Colors set this to 16 bits If you operate Remote Desktop with greater color depth e g 32 bit your windows may have a different appearance than they do on the instrument display because transparency becomes enabled 4 Click the Local Resources tab Selecting the Disk Drives checkbox enables you to transfer data between the remote desktop and the local ...

Страница 64: ...the X Series Signal Analyzers Remotely Step Notes 5 Click the Experience tab To Optimize the performance of the Remote Desktop session choose the appropriate connection format from the drop down list 64 Getting Started and Troubleshooting Guide ...

Страница 65: ...Notes 1 Click the X then click OK For full screen the X appears at the top center of the window For non full screen the X appears in a red box at the right of the window s title bar Or 2 When the remote desktop is full screen move the cursor to the bottom left of the window Click Start Disconnect Click Disconnect You are asked to confirm that you want to disconnect Getting Started and Troubleshoot...

Страница 66: ...r these versions or later IE Chrome Safari FireFox iOS Safari Chrome for Android 11 35 8 34 8 39 Accessing the instrument through the Internet It is possible to access and control the instrument through the Internet and World Wide Web or a local internet using the built in Embedded Server functionality This section provides details of how to use this functionality The instrument may also be access...

Страница 67: ...40b 00104 When the connection is made the welcome page appears 2 Click Control Instrument A password entry dialog appears By default this password is set at the factory as Keysight However you may subsequently change the password Press System I O Config Reset Web Password on the instrument front panel to change the password When the correct password has been entered the Control Instrument web page...

Страница 68: ...ep Notes Other Web Control Functions Selecting function in the navigation bar enables you to interact with the instrument through the web server SCPI Control This control enables you to send save recall and browse SCPI commands in the instrument 68 Getting Started and Troubleshooting Guide ...

Страница 69: ...s captured as a Portable Network Graphics PNG file to the default file name Screen png The image file can be saved to the client computer disk drive or copied to the Windows clipboard A typical screen capture image appears as follows Transfer Data This control enables you to get measurement and trace data that can be saved to your computer Getting Started and Troubleshooting Guide 69 ...

Страница 70: ...strument and provides access to Advanced options Access to Advanced options requires an additional login Advanced options enables you to Reset network Restore factory default network and Control Instrument configurations Windows Shortcuts and Miscellaneous Tasks This section provides a list of windows shortcuts key combinations that are useful when you operate the instrument without an attached mo...

Страница 71: ... Tools Embedded Web Server Using the X Series Signal Analyzers Remotely available in any Windows 7 system they are not commonly used when a mouse and keyboard are attached Getting Started and Troubleshooting Guide 71 ...

Страница 72: ...indows Start Menu Ctrl Esc Cycle through all open applications Alt Tab Select the first menu of a menu bar Alt Move through menu headings Left Arrow Right Arrow Open drop down a menu Down Arrow Move through items in an expanded menu Up Arrow Down Arrow Close the current menu selection Esc Cancel the current menu bar selection Alt Open an application s control menu usually the left most menu on the...

Страница 73: ...tes 1 Click Start Control Panel If not using a mouse press Ctrl Esc 2 Click Taskbar and Start Menu f you are not using a mouse use the shortcut key combinations specified in the Section Windows shortcuts key combinations on page 80 to make these selections 3 Click the Taskbar tab The Taskbar and Start Menu Properties dialog appears 4 Select the Auto hide the taskbar check box If you are not using ...

Страница 74: ...4 Using Window Tools Windows Shortcuts and Miscellaneous Tasks 74 Getting Started and Troubleshooting Guide ...

Страница 75: ... Windows 7 on page 78 Returning an Analyzer for Service on page 79 No operator serviceable parts inside Refer servicing to qualified personnel To prevent electrical shock do not remove covers Keysight X Series Signal Analyzer Getting Started and Troubleshooting Guide 75 ...

Страница 76: ...tion with Alt Tab Review the measurement procedures being performed when the problem first appeared Are all of the settings correct If the analyzer is not functioning as expected return the analyzer to a known state by pressing Mode Preset Some analyzer settings are not affected by a Preset If you wish to reset the analyzer settings press System Power On Restore Power On Defaults Is the measuremen...

Страница 77: ...will be necessary to perform the entire process again Is the analyzer displaying an error message If so refer to the Instrument Messages Guide Check if the external frequency reference is selected but not available Verify that it is selected by pressing Input Output Freq Ref Input If External is selected changing the setting to Sense allows the analyzer to sense the presence of an external referen...

Страница 78: ...rmance and measurement speed Those that can be safely modified are described in Settings that can be changed on page 44 The X Series if Signal Analyzers operate in an open Windows environment so you can install software on the instrument However installation of non approved software may affect instrument performance Keysight does not warrant the performance with non approved software installed 78 ...

Страница 79: ...eplacement parts contact the nearest Keysight Technologies office listed below In any correspondence or telephone conversations refer to your analyzer by its product number full serial number and software revision Press System Show System and the product number serial number and software revision information will be displayed on your analyzer screen A serial number label is also attached to the re...

Страница 80: ...77 1571 Belgium 32 0 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 0 10 855 2100 France 0825 010 700 0 125 Euros minute Germany 49 0 7031 464 6333 Ireland 1890 924 204 Israel 972 3 9288 504 544 Italy 39 02 92 60 8484 Netherlands 31 0 20 547 2111 Spain 34 91 631 3300 Sweden 0200 88 22 55 Switzerland 0800 80 53 53 United Kingdom 44 0 118 9276201 Other European Countries http www keysight com find c...

Страница 81: ...ur analyzer after the warranty has expired Call your Keysight Technologies office for full details If you want to service the analyzer yourself after the warranty expires you can download the service documentation that provides all necessary test and maintenance information from the Keysight web page Getting Started and Troubleshooting Guide 81 ...

Страница 82: ...g shipping container The carton must be both large enough and strong enough to accommodate the analyzer A double walled corrugated cardboard carton with 159 kg 350 lb bursting strength is adequate Allow at least 3 to 4 inches on all sides of the analyzer for packing material 3 Surround the equipment with three to four inches of packing material and prevent the equipment from moving in the carton I...

Страница 83: ...Kit Connector Kit Overview on page 84 Use of the Test Port Adaptors on Instrument RF Input 2 on page 85 Input Connector Vise Assembly on page 86 Keysight X Series Signal Analyzer Getting Started and Troubleshooting Guide 83 ...

Страница 84: ...e Used on RF Input 1 as a connector saver to connect 3 5 mm or sma devices or cables Adaptor 2 4 mm Female 2 4 mm Female Used on RF Input 1 as a connector saver 1 mm female test port to 1 mm female adaptor Used on RF Input 2 as a connector saver 1 mm female test port to 1 85 mm female adaptor Used on RF Input 2 as a connector saver and to connect 1 85 mm devices or cables Torque Wrench Double ende...

Страница 85: ...ctors are engaged will damage the input connector on the instrument 1 Align the test port adaptor with the 1 mm RF Input 2 connector on the instrument and engage the nut of the input connector with the adaptor Fully engage the threads but do not attempt to fully torque the connection 2 Locate the torque wrench in the kit Locate a 14 mm open end wrench 3 Place the jaws of the open end wrench onto t...

Страница 86: ...o 9 inch pounds 2 Loosen the Vise Clamp screw so that the vise jaws clear the body of the standard adaptor that is to be used 3 Align the flats of the 1 mm female end of the adaptor with the vise jaws and tighten the Vise Clamp screw until the jaws grip the adaptor Loosen the screw slightly to allow the adaptor to slide in the jaws 4 Align the adaptor with the 1 mm input connector on the instrumen...

Страница 87: ...rom damage when connections are made to the adaptor that exceed 4 inch pounds torque The Vise Clamp is only a rotation prevention device It will not protect the input connector against damage from bending due to heavy side to side or up down loading on the adaptor Getting Started and Troubleshooting Guide 87 ...

Страница 88: ...e 2 4 mm input connector on the instrument and engage the nut of the input connector with the adaptor Torque the bulkhead 2 4 mm input connector with an 8 mm or 5 16 torque wrench to 8 inch pounds while holding the adaptor with a 7 mm open end wrench Refer to Connector Care on page 89 88 Getting Started and Troubleshooting Guide ...

Страница 89: ...al Inspection on page 90 Connector Contacts on page 91 Making Connections on page 93 Cleaning Connectors on page 94 Replacement Parts on page 95 Keysight X Series Signal Analyzer Getting Started and Troubleshooting Guide 89 ...

Страница 90: ...damage that cannot be seen without magnification generally have no effect on electrical or mechanical performance Magnification is of great use in analyzing the nature and cause of damage and in cleaning connectors but it is not required for inspection Obvious defects or damage Examine the connectors first for obvious defects or damage Plating Bare metal showing Burrs or blisters Deformed threads ...

Страница 91: ... contacts Notice the location of the cross hairs in relationship to the center of the figures Contact integrity Refer to the following for visual guidelines when evaluating the contact integrity of a connector Concentricity The following examples show the concentricity of both the male and female 1 0 mm connectors Female connector Getting Started and Troubleshooting Guide 91 ...

Страница 92: ...7 Connector Care Connector Contacts Male connector 92 Getting Started and Troubleshooting Guide ...

Страница 93: ...em together As the center conductors mate there is usually a slight resistance 6 Do not twist one connector into the other like inserting a light bulb This happens when you turn the device body rather than the connector nut Major damage to the center conductor and the outer conductor can occur if the device body is twisted 7 Initial tightening can be done by hand or with an open end wrench Tighten...

Страница 94: ...a small cleaning swab do not apply alcohol directly to the parts 4 When using isopropyl alcohol to clean connectors do not allow the liquid to flow down inside the connector This may cause measurement errors due to residue inside the connector If possible keep the connector facing down 5 Gently wipe connecting surfaces with the end of the cleaning swab 6 Blow dry with compressed air 7 Inspect and ...

Страница 95: ...er Test Port Adaptor 1mm f to 1mm f Y1900B Test port Adaptor 1mm f to 1 85mm f Y1901B 2 4 mm f to 2 4 mm f 11900B 2 4 mm f to 2 92 mm f 11904B Wrench Torque Special Double end 14mm open end 4 and 10 lb in 8710 2819 Vise Kit N9041 60032 Getting Started and Troubleshooting Guide 95 ...

Страница 96: ...This information is subject to change without notice Keysight Technologies 2017 Edition 1 July 2017 N9041 90001 www keysight com ...