6 Keysight Connector Kit
Input Connector Vise Assembly
Input Connector Vise Assembly
Keysight Technologies recommends the use of the supplied Keysight Technologies
test port adaptors when using RF Input 2, the 1 mm input connector. The test port
adaptors attach to the external threads on the instrument bulkhead input connector.
However, if Keysight Technologies standard adaptors will be used instead of the test
port adaptors, the use of the supplied Connector Vise assembly is recommended to
prevent damage to the 1 mm input connector due to over torque.
Standard adaptors other that those manufactured by Keysight Technologies may
not work with the Vise Clamp.
USE CAUTION WHEN INSTALLING ADAPTORS! Rotation of the adaptor after the
connectors are engaged will damage the input connector on the instrument.
To use Keysight Technologies standard adaptors with the Vise Clamp:
1. Install the Vise Clamp with the two supplied pan-head screws. Tighten the
screws to 9 inch-pounds.
2. Loosen
the
Vise
Clamp
screw
so
that
the
vise
jaws
clear
the
body
of
the
standard
adaptor
that
is
to
be
used.
3. Align the flats of the 1 mm female end of the adaptor with the vise jaws and
tighten the Vise Clamp screw until the jaws grip the adaptor. Loosen the screw
slightly to allow the adaptor to slide in the jaws.
4. Align the adaptor with the 1 mm input connector on the instrument and engage
the nut of the input connector with the adaptor. Fully engage the threads, but do
not attempt to fully torque the connection.
5. Re-align
the
jaws
of
the
vise
so
they
make
good
contact
with
the
flats
on
the
adapt
o
r.
Do
this
by
pulling
back
on
the
vise
jaws
while
tightening
the
vise
clamp
screw.
DO
NOT
OVER
TIGHTEN.
6. Tighten the input connector nut to 4 inch-pounds using the 4-inch-pounds end of
the 14 mm torque wrench.
7. Re-tighten the vise clamp screw to snugly lock the jaws against the flats of the
adaptor. DO NOT OVER TIGHTEN.
8. If the standard adaptor will remain on the input connector for a long period of
time, periodically check the Vise Clamp screw to ensure that it remains snugly
tightened.
86
Getting Started and Troubleshooting Guide
Содержание x-series
Страница 6: ...6 Getting Started and Troubleshooting Guide ...
Страница 18: ...1 Quick Start Instrument Information 18 Getting Started and Troubleshooting Guide ...
Страница 56: ...3 Instrument Operating System Disk Drive Recovery Process 56 Getting Started and Troubleshooting Guide ...
Страница 74: ...4 Using Window Tools Windows Shortcuts and Miscellaneous Tasks 74 Getting Started and Troubleshooting Guide ...
Страница 92: ...7 Connector Care Connector Contacts Male connector 92 Getting Started and Troubleshooting Guide ...