![Keysight U8903B Скачать руководство пользователя страница 384](http://html.mh-extra.com/html/keysight/u8903b/u8903b_user-manual_1961457384.webp)
7
Test Sequence Application
384
Keysight U8903B User’s Guide
Deemphasis
Press the
Deemphasis
softkey to select the de-emphasis filter.
– None
– 50
μ
s
– 75
μ
s
– Custom
Notch Filter
Press the
Notch Filter
softkey to configure the notch filter settings.
– HP8903B
Enable or disable the HP8903B notch filter mode.
– Custom Notch
Enable or disable custom notch filter.
– Center Freq
Set the center frequency value. This is only available when the Custom Notch is enabled.
– Bandwidth
Set the bandwidth value. This is only available when the Custom Notch is enabled.
Average Points
Press the
Average Points
softkey to set the number of measurement readings to be used for the average calculation. This
is useful for noisy signals. Applying the average points will smooth out the fluctuations introduced by the noise that causes
the inconsistencies in the measurement reading.
Sample Size
Press the
Sample Size
softkey to select the number of samples to be acquired for the measurement.
– 2048
– 4096
– 8192
– 16384
– 32768
– 65536
– 131072
– 262144
– 524288
– 1M
– 2M
Input Range
Press the
Input Range
softkey to configure the input range settings.
– Auto Range
Enable or disable the auto input range.
If auto range is enabled, each analog channel input range is determined automatically, based on the level of the input
signal. If the input signal level changes beyond the ranging threshold, auto range will cause the input ranging circuits to
move up or down for a proper ranging.
If auto range is disabled, you can set a fixed input voltage range for each analog input channel.
– Track 1st Ch
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to
channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual
channels.
– Voltage Range
Set the input voltage range.
Table 7-53
Receiver Sensitivity > Signal Analysis settings menu description (continued)
Menu
Description
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Страница 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 42: ...1 Getting Started 42 Keysight U8903B User s Guide LCD display Figure 1 4 U8903B LCD display 2 3 1 4 ...
Страница 96: ...2 Operation and Features 96 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 406: ...7 Test Sequence Application 406 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...9 Characteristics and Specifications 418 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 574: ...A Appendixes 574 Keysight U8903B User s Guide Figure A 11 Bluetooth generator sine waveform generation ...