Operation and Features
2
Keysight U8030A Series User’s Guide
57
Clear the overcurrent condition
The OCP trip can only occur if the corresponding output is enabled. When the
OCP trip occurs, the corresponding output is disabled automatically.
– If the OCP trip occurs on Output 1, the display shows
triP
on line 1 and the
OCP1
annunciator blinks.
– If the OCP trip occurs on Output 2, the display shows
triP
on line 2 and the
OCP2
annunciator blinks.
1 Select the output channel with the OCP trip.
– Press
[1]
to select Output 1, or
– Press
[2]
to select Output 2.
– The corresponding annunciator (
OUT1
or
OUT2
) turns on.
2 Clear the OCP trip.
– Press the
[Over Current]
button again to clear the OCP trip.
NOTE
If the OCP trip persists, decrease the current limit settings to clear the tripping.