Calibration Procedures
1
Keysight U3606B Service Guide
19
Adjustment is recommended
Specifications are only guaranteed within the period stated from the last
adjustment. Keysight recommends that re-adjustment should be performed
during the calibration process for best performance. This will ensure that the
instrument will remain within the specifications for the next calibration interval.
This criterion for the re-adjustment provides the best long-term stability.
Performance data are measured during the
but this does not
guarantee that the instrument will remain within these limits unless the
adjustments are performed.
Refer to the
on page 100 and verify that all the adjustments
have been performed.
Time required for calibration
The instrument can be automatically calibrated under computer control. With
computer control you can perform the complete calibration procedure and
performance verification tests once the instrument is warmed-up (see