1
Calibration Procedures
100
Keysight U3606B Service Guide
Calibration Count
The calibration count feature provides an independent “serialization” of your
calibrations. You can query the instrument to determine how many calibrations
have been performed.
1
In adjustment mode, press
[MinMax]
to view the calibration count.
2
Take note of the calibration count to keep track of the number of calibrations
that have been performed.
3
The instrument will return to the calibration operation after a few seconds of
inactivity.
NOTE
– The instrument has been calibrated before it left the factory. You are
recommended to record the initial value of the calibration count once you
receive the instrument.
– The calibration count increments up to a maximum of 32767 after which it
rolls over to 0. Since the value increments by one for each calibration point,
a complete calibration may increase the value by many counts.
– The calibration count is stored in the nonvolatile memory, and does not
change when power has been turned off, after a Factory Reset (
*RST
command), or after an Instrument Preset (
SYSTem:PRESet
command).