Calibration Procedures
86
Keysight U3606B Service Guide
AC current gain adjustment procedure
Review the
on page 25 and
on page 77 sections before beginning this procedure.
1
Press
[
]
twice to enter the AC current gain calibration. The AC annunciator
will turn on according to your selection.
2
The display shows the uncalibrated value and the reference value of the
adjustment item.
3
Configure each adjustment item shown in
.
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input.
7
Press
to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
– Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
– An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the AC current gain adjustments using the
“AC current verification test”
I
NOTE
Always complete tests in the same order as shown in
OUT
SBY
Содержание U3606B
Страница 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Страница 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 104: ...1 Calibration Procedures 104 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...