Calibration Procedures
6
Adjustments procedures
U3606A User’s and Service Guide
203
7
Press
to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
•
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
•
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the frequency gain adjustments using the
verification test” on page 161
.
Resistance gain adjustment procedure
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
This procedure adjusts the gain for the two- wire resistance function. The
gain for the 100 M
Ω
range is derived from the 10 M
Ω
range and does not
have a separate adjustment point.
1
Press
to enter the resistance gain calibration.
2
The primary display will show the uncalibrated value and the upper
secondary display will show the reference value of the adjustment item.
3
Configure each adjustment item shown in
Table 6-5
Frequency gain adjustment
Adjustment item
Input frequency
Input voltage (V
rms
)
1000.00 Hz, 1 V
rms
1 kHz
1 V
rms
O U T
S B Y
N O T E
If the zero offset adjustment procedure has been recently performed prior
to the resistance gain calibration procedure, the adjustment items “Short”
and “OPEn” can be omitted.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...