Verification and Performance Tests
5
Performance Verification Tests
U3606A User’s and Service Guide
161
Frequency gain verification test
1
Connect the calibrator to the front panel
V
(red) and
LO
(black) input
).
2
Press
Hz ms %
to select the frequency function.
3
Select each range in the order shown below. Provide the indicated input
frequency and voltage. Compare the measurement results to the
appropriate test limits shown in
. (Be certain to allow for
appropriate source settling when using the Fluke 5520A.)
10 M
Ω
10 M
Ω
±0.0255 M
Ω
10 M
Ω
100 M
Ω
±0.205 M
Ω
[1] Specifications stated are for 2-wire resistance measurements using the Null math operation.
Without Null, add 0.2
Ω
additional error.
Table 5-8
Frequency gain verification test
Input frequency
Input voltage
Range
Error from
nominal 1 year
1 kHz
1 V
100 kHz
±0.0032 kHz
Table 5-7
Resistance gain verification test (continued)
Input resistance
Range
Error from nominal 1 year
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...