Performance Tests and Calibration
7
Keysight U1401B User’s and Service Guide
143
Performance Verification Tests
Self-verification
To perform a self-verification on the output voltage level of the instrument:
1
Turn the rotary switch to the
/
position.
2
Short the input test leads for voltage measurement, then press
momentarily to zero the residual of thermal effect until the measurement value
is stable.
3
Connect the positive ends of input and output together.
4
Connect the negative ends of input and output together.
5
Set output value to
+
4.5000 V.
6
Observe the measurement value in the primary display.
for functions that can be self-verified.
is for reference only. Refer to
Chapter 8, “Characteristics and
on page 157 for the detailed specifications.
Table 7-2
Functions that can be self-verified
Rotary switch position
Output value
Measuring value (input)
/
+4.5000 V
DC 4.5000 V
/
+25.0000 mA
DC 25.0000 mA
/
100 Hz
100.00 Hz
0.39~99.60%
0.3~99.6%
±5 V
AC 4.9586 V
±12 V
AC 11.959 V
Содержание U1401B
Страница 1: ...Keysight U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 61: ...Making Measurements 3 Keysight U1401B User s and Service Guide 61 Figure 3 1 DC voltage measurement ...
Страница 76: ...3 Making Measurements 76 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 107: ...Application Examples 5 Keysight U1401B User s and Service Guide 107 Figure 5 5 Zener diode test ...
Страница 128: ...5 Application Examples 128 Keysight U1401B User s and Service Guide Figure 5 17 Verifying a frequency transmitter ...