Keysight PD1000A Power Device Measurement System Startup Guide
51
Running the IV and CV Tests
IV and CV Device Characterization
-- For the H71 option, use the UHCU named:
SMUn:UHC1500
.
11
Input – HVSMU:
Enter the name of the HVSMU used for input (for example,
SMU3:HV
).
–
Use with the B150xA option H21 (20 A) only.
–
Leave this field blank for options H51 and H71.
–
Look up the module name from the HVSMU input field in the
Module
Selector
tab of the Configuration dialog in EasyEXPERT. (
See “Configure
12
Input – SMU:
Enter the name of SMU used for SMU input (for example,
SMU3:MF).
–
Use with the B150xA option H21 (20 A) only.
–
Leave this field blank for options H51 and H71.
–
Look up the module name from the HVSMU input field in the
Module
Selector
tab of the Configuration dialog in EasyEXPERT. (
See “Configure
13
Input – Gate SMU:
Enter the Gate SMU name:
–
For the B1506A: always use
SMU2:MC
.
–
For the B1505A:
-- Identify the SMU that connects to the gate of the DUT.
-- Look up the SMU/module name from the Modules tab of the
Configuration dialog in EasyEXPERT. (
page 37.)
Temperature Settings:
1
Hot Plate VISA Address:
From the appendix, “Configure the InTEST Thermal
Platform Controller,” starting on page 95.
2
Settling Temperature Timeout (in minutes):
The maximum time that you allow for
the Hot Plate to reach the specified temperature.
3
Temperature Range (Radius) (°C):
The acceptable tolerance of the measured
temperature.
4
Transistor Temperature Delay (S):
The time to wait (in seconds) after the target
temperature is reached.
5
Set the Hot Plate to room temperature after measurement:
Check the box if you
want the Hot Plate reset to ~23
°
C after temperature measurements
complete. Leave the box unchecked to retain the existing temperature.
6
Maximum Hot Plate Temperature Allowed:
Specify the maximum temperature for
the Hot Plate (
°
C).
Содержание PD1000A
Страница 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Страница 2: ......
Страница 4: ......
Страница 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Страница 96: ...94 Keysight PD1000A Power Device Measurement System Startup Guide PD1000A System Specifications ...
Страница 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Страница 107: ......