3-6 Keysight N9923A Service Guide
Tests and Adjustments
Operational Check
3-
Operational Check
1.
Preset the instrument, press Preset (.) > Preset.
2.
Leave the PORT 1 and Port 2 connectors open.
3.
Press: Scale/Amptd > Ref Pos > 2 > Enter
4.
Press: Scale > 1 > Enter
5.
The trace should be between -1 and +1 dB. (S11 is displayed.)
6.
Press: Measure
7.
If the S22 soft key exists, press it to display S22.
a.
The trace should be between -1 and +1 dB.
If the operational check fails, allow the instrument to warm up for 15
minutes and try again.