Installation Note N9032-90004

23

Option EXM, External Mixing Upgrade Kit

Utilities, Adjustments, and Performance Verification Tests

Calibration Software and specified test equipment is required to perform the adjustments, and can 

be used to automate the performance verification testing. Information on how to obtain this 

software can be found at:

http://www.keysight.com/find/calibrationsoftware

Utilities Required

None

Adjustments Required

The following adjustments are the minimum set required to ensure that the newly installed 

hardware is functioning properly

— IF Input Gain Adjustment
— LO Output Power Adjustment

Performance Testing Required

The following performance verification tests are the minimum set required to ensure that this newly 

installed option is functioning properly.

— IF Input Gain Accuracy
— LO Output Power Accuracy

A full calibration is required to assure the instrument meets all specifications

The end user must ultimately determine whether they want a full calibration to be performed after 

the installation of this upgrade or not. If a full calibration is required, arrangements regarding the 

level of calibration must be made between the end user and the calibration provider.
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your 

local Keysight office access the following URL, or if in the United States, call the following 

telephone number:

http://www.keysight.com/find/assist

1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)

Содержание N9032B PXA

Страница 1: ...Option EXM External Mixing Upgrade Kit Keysight N9032B PXA Signal Analyzer Installation Note ...

Страница 2: ...to the implied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Manual Part Number N9032 90004 Edition Edition 2 October 2021 Printed in USA Malaysia Published by Keysight Technologies Inc 1400 Fount...

Страница 3: ...s to upgrade a N9032B PXA signal analyzer to add option EXM External Mixing This kit includes parts to upgrade a PXA regardless of frequency range option installed In addition to installing the hardware to support option EXM a license for the following option will also be installed N9032B EXM External Mixing Products Affected N9032B PXA Serial Numbers All To Be Performed By X Keysight Factory Keys...

Страница 4: ... Lock Internal Tooth inch 2190 0067 1 Nut Hex Double Chamfer 1 4 36 THD 125 IN THK Stainless Steel 2950 0223 1 Cable Assembly External Mixing Front Panel W30 N9020 20166 1 Cable Assembly A13J8 to W30 W29 N9020 20167 1 Label Warning for Top Brace N9030 80018 17 Screw M3 x 0 5 6 mm long flat head 0515 1946 1 Entitlement Certificate 5964 5178 1 Entitlement Certificate Envelope 5967 7169 1 Installatio...

Страница 5: ...t and display the measurement screen The instrument will probably display a spectrum analyzer screen and you will see the instrument sweeping There should be no alignment failures If there are failures investigate and fix the problem before continuing Before you disassemble the instrument turn the power switch to Standby After the instrument has completely shut down unplug the instrument Failure t...

Страница 6: ...instrument four bottom feet 4 Remove the four instrument bottom feet 6 5 Using the T 20 driver remove the four screws including washers 3 that hold the rear feet 4 in place 6 Pull the instrument cover 5 off towards the rear of the instrument Figure 1 Instrument Outer Case Removal If the instrument is placed on its face during any of the following procedures be sure to use a soft surface or soft cl...

Страница 7: ...efer to Figure 2 To remove the top brace 1 use the T 10 driver to remove the twelve screws 2 0515 0372 attaching the top brace to the chassis Remove the wire hold down as shown in Figure 3 Remove and discard the seventeen screws 3 0515 1946 attaching the top brace Figure 2 Top Brace Removal ...

Страница 8: ...8 Installation Note N9032 90004 Option EXM External Mixing Upgrade Kit Figure 3 Wire Hold Down ...

Страница 9: ...er to Figure 4 Using the T 10 driver remove the twelve screws 1 0515 1035 six on each side to detach the Front Frame Assembly from the chassis 3 Refer to Figure 5 Pull the Front Frame Assembly carefully away from the chassis Remove the ribbon cable W1 from the motherboard The cable has locking tabs on each side pinch and pull to release Figure 4 Front Frame Removal Make sure any connectors on the ...

Страница 10: ...10 Installation Note N9032 90004 Option EXM External Mixing Upgrade Kit Figure 5 Front Panel Cable ...

Страница 11: ...ed on the front panel EXT mixer connector 3 Locate semi rigid coax cable part number N9020 20167 in the upgrade kit This is W29 Connect the end with the SMA male connector A13A1J8 with the SMA female connector pointing towards where the front panel would be The long straight section of W29 should be parallel to the casting of the A13 Front End and level Refer to Figure 6 Torque the cable nut to 10...

Страница 12: ...ge with the recess in the front frame assembly Refer to Figure 7 Secure the connector using the lockwasher and hex nut from the rear Torque to 21 inch pounds Figure 7 Attaching SMA Female Connector 4 Locate the SMA termination removed on page 11 step 2 Attach this SMA termination to the front panel EXT Mixer SMA connector 5 Locate the External Mixing Front Panel semi rigid coax assembly in the kit...

Страница 13: ...Installation Note N9032 90004 13 Option EXM External Mixing Upgrade Kit Figure 8 Connecting W30 to Ext Mixer Connector Figure 9 Proper Orientation of W30 ...

Страница 14: ... Figure 10 Reattach the ribbon cable W1 Ensure the locking tabs are engaged 2 Refer to Figure 11 Carefully position the Front Frame Assembly onto the chassis Ensure no cables are crushed Replace the twelve screws 1 0515 1035 six on each side of the chassis Torque to 9 inch pounds Figure 10 Front Panel Cable ...

Страница 15: ...Installation Note N9032 90004 15 Option EXM External Mixing Upgrade Kit Figure 11 Front Frame Replacement ...

Страница 16: ... panel is installed cables W29 and W30 typically do not align perfectly This is OK since the cables are slightly flexible If additional alignment is needed loosen W29 at A13J8 and re position 4 Refer to Figure 13 Align and connect cables W29 and W30 Hand tighten the nut Figure 12 W29 and W30 Before Alignment ...

Страница 17: ...ing Upgrade Kit Figure 13 W29 and W30 Properly Aligned 5 Refer to Figure 14 Use a open end wrench to prevent the SMA female connector on W29 from twisting Use a 5 16 torque wrench to torque the nut on W30 to 10 inch pounds Figure 14 Torque Cable W30 onto W29 ...

Страница 18: ...replace the top brace place the brace in the correct position and attach to the chassis using the twelve screws 2 0515 0372 Attach the top brace using the seventeen screws 3 0515 1946 included in the kit 2 Refer to Figure 16 Reinstall wire hold down 1 3 Torque all screws to 9 inch pounds Figure 15 Top Brace Replacement ...

Страница 19: ...Option EXM External Mixing Upgrade Kit Figure 16 Wire Hold Downs 4 In the upgrade kit locate the Warning Label N9030 80018 5 Refer to Figure 17 Attach the Warning Label to the top brace as shown Figure 17 Add Warning Label to Top Brace ...

Страница 20: ...o the gasket groove in the Front Frame Assembly 2 Replace the four rear feet to the rear of the instrument Torque the rear feet screws 0515 1619 and 3050 0893 washers to 21 inch pounds 3 Replace the bottom feet by sliding into place until they snap in Install the locks by pressing in flat 4 Replace the handle straps on both sides of the instrument Torque the handle strap screws to 21 inch pounds F...

Страница 21: ...m a virus scan on this device before use 5 Save the License File to the root directory of the USB Storage Device 6 Connect the USB Storage Device to one of the analyzer s USB ports Connect a mouse to another USB port Windows will detect the new hardware and may display the configuration menu shown in Figure 19 This menu may be configured according to your preferences Figure 19 USB Storage Device C...

Страница 22: ...ust be restarted Press File Exit An Exit Analyzer dialog box will appear press Enter to confirm the exit 2 Double click on the LaunchXSA icon on the Windows desktop Wait for the XSA application to finish starting the analyzer should be sweeping 3 Press System Show System on the analyzer to display a list of all displayed options You should see the following option listed N9032B EXM External Mixing...

Страница 23: ...ed The following performance verification tests are the minimum set required to ensure that this newly installed option is functioning properly IF Input Gain Accuracy LO Output Power Accuracy A full calibration is required to assure the instrument meets all specifications The end user must ultimately determine whether they want a full calibration to be performed after the installation of this upgr...

Страница 24: ...This information is subject to change without notice Keysight Technologies 2021 Edition 2 October 2021 N9032 90004 www keysight com ...