Installation Note N9032-90004
11
Option EXM, External Mixing Upgrade Kit
Add a Cable to the A13 Front End Assembly J8
1.
Remove the SMA termination on A13A1J8. A13A1J8 is to the left of A13A1J13.
2.
Save this SMA termination. It will be installed on the front panel EXT mixer connector.
3.
Locate semi-rigid coax cable, part number
, in the upgrade kit. This is W29.
Connect the end with the SMA male connector A13A1J8, with the SMA female connector
pointing towards where the front panel would be. The long, straight section of W29 should be
parallel to the casting of the A13 Front End and level. Refer to
. Torque the cable nut to
10 inch-pounds.
Figure 6
Orientation of W29
Содержание N9032B PXA
Страница 1: ...Option EXM External Mixing Upgrade Kit Keysight N9032B PXA Signal Analyzer Installation Note ...
Страница 8: ...8 Installation Note N9032 90004 Option EXM External Mixing Upgrade Kit Figure 3 Wire Hold Down ...
Страница 10: ...10 Installation Note N9032 90004 Option EXM External Mixing Upgrade Kit Figure 5 Front Panel Cable ...
Страница 15: ...Installation Note N9032 90004 15 Option EXM External Mixing Upgrade Kit Figure 11 Front Frame Replacement ...